Содержание KM-CAL-907

Страница 1: ......

Страница 2: ...ssure measuring pressure Calibrating pressure voltage transmitter Calibrating 2wire pressure transmitter Calibrating pressure switch Others features z Manual step source and auto step and sweeping step source z Measurement source mA display z Measurement wave filter function z Measurement manual holding function z Pressure source auto holding function 1 Contact Us To purchase parts obtain operatio...

Страница 3: ...se contact the vendor from which you purchased the product as soon as possible Refer to the replacing part list in 15 3 in the Manual if you want to order the replacing parts A set of Industrial testing Lead CL727220 A set of Testing Lead Tp727110 A set of Alligator clip CC807130 A quick reference guide A User s Manual One Fuse 50mA 250V One Fuse 63mA 250V ...

Страница 4: ...e equipment under test Refer to Table 1 for the explanation of the international electric symbols adopted by the calibrator or the user s manual Table 1 Explanations of International Electrical Symbols EARTH GROUND WARNING INFORMATION Warning To avoid possible electric shock or personal injury z Do not apply more than the rated voltage as marked on the calibrator between terminals or between any t...

Страница 5: ...ing the meter z When using the probes keep fingers behind the finger guards on the probes z Connect the common test lead before connecting the live test lead When disconnecting test leads disconnect the live test lead first z Do not use the meter if it operates abnormally Protection may be impaired When in doubt have the meter inspect z Do not operate this instrument in areas where inflammable or ...

Страница 6: ...cified replacement parts z To avoid false reading which could lead to possible electric shock or personal injury replace the batteries as soon as the low battery indicator appears z Caution To avoid possible damage to meter or to equipment under test z Disconnect the power and discharge all high voltage capacitors before testing continuity z Use the proper jacks functions and ranges for the measur...

Страница 7: ...7 4 Familiar With the Calibrator Figure 1 Entire Graph Figure 2 Measurement Source Terminals ...

Страница 8: ...ble 2 Measurement Source Terminals Figure 3 keys Terminal Function 1 Measurement Signals DCmV 2 Measurement Signals DCV FREQ 3 Measurement Signals DCmA 4 All the common return terminals of measurement function 5 Source Signals DCmA Source Signals Simulate mA 6 Source Signals DCmA LOOP Terminal 24VDC Loop Power Terminal 7 All the common return terminals of source function 8 Source Signals DCV FREQ ...

Страница 9: ...ction 13 Power key Turn on or off the power 14 Backlight Turn on or off the backlight 15 V Key Select measurement source DCV function 16 PULSE Key Select source pulse function 17 mA Key Select measurement source mA function 18 FREQ Key Select measurement source FREQ function 19 Key Select measurement continuity function Select calibration pressure switch function 20 RANGE Key Select measurement so...

Страница 10: ...uto stepping or sweeping function Measured value holding Unlock the pressure switch state 25 Key In DCmA source function select the auto wave type In frequency or pulse source set the frequency value 26 25 100 Key In mA source function select 25 or 100 manual step output type In pulse number frequency source set the amplitude value Measuring average value Display Screen Figure 4 shows a typical di...

Страница 11: ...wer Supply on l Measurement pressure m Beeper of measurement continuity n Pressure reading zero off o Measured value holding p Measuring average value q Unit of pressure r Main display date and message s Present Unit of main display date Ω t DCmA 25 or 100 manual step source u Start source number of pulses DCmA auto stepping or sweeping source function v 24V Loop Power Supply on Figure 4 typical L...

Страница 12: ... which you purchased the instrument for a service of inspecting or adjusting the internal assembly z In case of failure Should the instrument begin to emit smoke give off an unusual odor or show any other anomaly immediately turn off the POWER key If you are using an Charger disconnect the plug from the wall outlet Also cut off power to the object under test that is connected to the input terminal...

Страница 13: ...r a prolonged period Be careful not to let a soldering iron or any other heat emitting object come into contact with the operation panel as the panel is made of thermoplastic resin z Before cleaning the instrument s case or operation panel disconnect the power cord plug from the wall outlet if you are using an Charger Use a soft clean cloth soaked in water and tightly squeezed to gently wipe the o...

Страница 14: ...to any heat source z Exposed to frequent mechanical vibration z Close to any noise source such as high voltage equipment or motive power sources z Close to any source of intensive electric or electromagnetic fields z Exposed to large amounts of greasy fumes hot steam dust or corrosive gases z Exposed to unstable or a risk of explosion due to the presence of flammable gases Note z Use the instrumen...

Страница 15: ...he given ambient temperature for at least one hour to ensure that the instrument is free from condensation before using the instrument Installing or Replacing the Batteries Warning To avoid electrical shock always remove the source or measurement lead cables from the object under test as well as from the instrument itself before opening battery door Caution z To avoid the risk of fluid leakage or ...

Страница 16: ...e type in the battery holder with their positive and negative electrodes positioned correctly as indicated on the holder Step 4 After replacement reattach the battery holder cover Indication of Battery Level The battery replacement indicator shows the battery level in five steps according to the measured voltage of the batteries Full battery The battery level is below 50 full The battery level is ...

Страница 17: ...z Do not use any Charger other than the dedicated Charger from the Company z Do not charge non Ni Cd Ni MH batteries or wasted batteries Step 1 Make sure the calibrator is turned off Step 2 Insert the plug of the optional Charger into the Charger connection jack Note z Turn off the calibrator before connecting or disconnecting the Charger from AC power plugging in out the Charger connection jack z...

Страница 18: ... LCD can be backlit Pressing the key turns on the backlight while pressing the key once again turns it off This feature makes it easier for you to view the LCD when operating the calibrator in dark places or when carrying out source or measurement Battery life shortens when the calibrator is operated on batteries Note The backlight automatically turns off after 10 seconds Press the key once more t...

Страница 19: ...Caution z Do not apply any voltage to the output terminals for ranges other than 4 20mA simulating transmitter output Otherwise the internal circuitry may be damaged 7 1 Connecting Cables to Terminals For DC voltage frequency pulse Figure 5 Step 1 Connect the black lead cable for source to the COM output terminal and the red lead cable to the VHz output terminal Step 2 Connect the other ends of th...

Страница 20: ...unction select the desired range from 1000mV 10V by pressing the RANG key The default value and unit of the selected source function and Figure 6 Sourcing DC Current range shall be displayed on the LCD Step 2 Set the output value digit by digit using S T keys Each pair of S T keys corresponds to each digit of the LCD reading Each press of the S T key increases or decreases the digit Increasing the...

Страница 21: ...ey once again The OFF appears on the LCD and no signals sourced between the terminals 7 3 Sourcing DC Current Step 1 Using the mA key to select the desired source function 20mA The default value and unit of the selected source function shall be displayed on the LCD Step 2 Set the output value digit by digit using S T keys Each pair of S T keys corresponds to each digit of the LCD reading Each pres...

Страница 22: ...20 mA Function You can set the source value in 4 mA or 16mA increments or decrements within 4 20 mA current Step 1 In DC current function press the 25 100 key to display 25 SET on the lower part of the screen and press once again to display 100 SET The default source value will be showed simultaneously Step 2 Using each pair of S T output setting keys set the signal in a step by step manner In 25 ...

Страница 23: ...g mode or in auto sweeping mode It requires 80 seconds to finish a 4 20mA cycle for auto sweeping mode and 20 seconds for auto stepping mode Step 1 In DC current function press key to display auto stepping mode signal on the lower part of the screen and press once again to display auto sweeping mode signal The default source value will be showed simultaneously Step 2 Pressing the ON key causes the...

Страница 24: ...rminals Tips z Press the START key again to continue the auto stepping and auto sweeping mode after stopping them and RUN mark displays on the lower part of the screen z Using the START key to start mA auto stepping and auto sweeping mode is only available when the source function is in ON state 7 3 3 mA display In mA source function the preset value displayed on main districts of the LCD And mA d...

Страница 25: ...rement set for mA can set the mA value only 7 3 4 4 20 mA simulating transmitter source Connect the calibrator and the loop power as listed in Figure 7 and operate in steps shown in sourcing DC current Figure 7 4 20 mA simulating transmitter source ...

Страница 26: ...p 3 Set the output value digit by digit using each pair of S T output setting keys Each pair of S T keys corresponds to each digit of the LCD reading Each press of the S T key increases or decreases the digit Increasing the digit from 9 or decreasing it from 0 causes the digit to overflow or underflow allowing you to set the output value without interruption Holding down the S T key continuously c...

Страница 27: ...ing down the S T key continuously changes the digit And the value won t change if it is increased or decreased to the Maxim or Minimum value Step 6 To re enter into the frequency set mode press the FREQ key to set the frequency Step 7 Pressing the ON key causes the SOURCE indicator on the LCD to change from OFF to ON The calibrator sources constant pulse signals responding to the preset frequency ...

Страница 28: ...GE key select a desired frequency range from 100Hz 1KHz 10KHz The selected function and the default range source value and unit shall be shown on the LCD Step 3 Set the output value digit by digit using each pair of S T output setting keys Each pair of S T keys corresponds to each digit of the LCD reading Each press of the S T key increases or decreases the digit Increasing the digit from 9 or dec...

Страница 29: ...git And the value won t change if it is increased or decreased to the Maxim or Minimum value Step 6 Pressing the CYC key enter into the pulse number set mode and the LCD shows the default number of 1 CYC in the part Step 7 Set the output value digit by digit using each pair of S T output setting keys Each pair of S T keys corresponds to each digit of the LCD reading Each press of the S T key incre...

Страница 30: ... is complete the calibrator automatically turns off the output and ceases operation The RUN symbol disappears from the LCD Step 12 To turn off the output press the ON key once again The OFF appears on the LCD and no signals sourced between the terminals Tips z The frequency range of the pulse could only be changed by pressing RANGE key in the frequency set mode z In the pulse sourcing process pres...

Страница 31: ...ty Warning z In an application where the calibrator is used together with the supplied lead cables for measurement the allowable voltage to ground of the input terminals is 60 Vpk maximum To avoid electrical shock do NOT use the calibrator at any voltage exceeding the maximum voltage to ground Tips z With the HOLD key you can hold the measured value z The reading of a measured value is updated dif...

Страница 32: ...igure 8 Measuring DC voltage mV continuity For DC voltage V Frequency measurement Figure 9 Step 1 Connect the black lead cable for measurement to the COM input terminal and the red lead cable to the VHz input terminal Step 2 Connect the other end of the cable to the measuring terminals of equipment under test while making sure the polarities are correct For DC current signal Figure 10 Step 1 Conne...

Страница 33: ...for the mA current input Figure 10 Measuring DC current z terminal and vice versa when wiring is extremely dangerous NEVER make this mistake z The current input terminals are equipped with a built in current input protection fuse Over current input to the terminals will cause the fuse to blow If the fuse is blown replace it with one with the specified ratings For details on fuse replacement see re...

Страница 34: ...the RANGE key select a desired range from 200mV 5V 50V The selected function and the measured value and unit shall be shown on the LCD 8 3 Measuring DC Current Step 1 Make sure the lead cables for measurement are not connected to the measuring instrument under test Step 2 Using the mA select DC Current measurement function Step 3 Connect the lead cables for measurement to the measuring terminals o...

Страница 35: ...ibrator as a loop power supply Figure 11 Using 24V loop power circuit to calibrate a 2 wire converter by undertaking the following steps Step 1 When the calibrator is in current measurement function pressing the LOOP key causes the LCD shows LOOP symbol And the built in 24V loop power of the calibrator will be turned on Step 2 Connect the calibrator with the loop current terminal of the supply con...

Страница 36: ...NGE key select the suitable range from 500Hz 5KHz 50KHz The selected function and the measured value and unit shall be shown on the LCD 8 5 Measuring Continuity Continuity measurement is used to detect the intactness of the circuit e g a resistance lower than 200 Using the key select continuity measurement function LCD displays continuity symbol on the LCD Connecting the devices as shown in Figure...

Страница 37: ...AVG key cancels the calculation and the AVG symbol disappears 8 7 Measured Value holding function Apart from the continuity measurement functions the reading hold function can be used to preserve the current measured value on the upper part of LCD which consequently doesn t refresh the measured value Pressing the HOLD key selects reading hold mode and LCD displays HOLD symbol To cancel the selecti...

Страница 38: ... avoid electrical shock do not use the calibrator at any voltage exceeding the maximum voltage to ground Tips z With the HOLD key you can hold the measured value z Under calibrating pressure transmitter mode 1 The reading of a measured value is updated differently responding to different measurement function LCD shows on the upper part when shifting the range If the input is over range the measure...

Страница 39: ...system do shut off the valve to release the pressure gradually before connecting the pressure module with the pipe Caution z To avoid any mechanical damage to the pressure module do not apply any force higher than 13 5 Nm 10 ft 1bs to the pressure pipe mouths or the module and the pipe mouth Do apply the specified force when connecting the pipe or the adapter z To avoid any damage to the pressure ...

Страница 40: ... upper part Step 3 Pressing the ON key the calibrator connects and senses the type of the pressure module and sets the range automatically If it fails to connect the LCD shows NO OP in the upper part Step 4 Zero off the reading following the pressure module manual When the reading overtops the 5 percentage of maxim value of the range the LCD shows ERR in the lower part Pressing the ZERO key initia...

Страница 41: ...brate Pressure Voltage Transmitter Calibrator can measuring voltage signal from transmitter under pressure mode Step 1 Connect the pressure module and calibrator as shown in Figure 13 Step 2 Pressing the ON key the calibrator connects and senses the type of the pressure module and sets the range automatically If it fails to connect the LCD shows NO OP in the upper part Step 3 Press the V key to se...

Страница 42: ...ssure value displayed on the LCD The LCD shows the present voltage value from the transmitter in the assistance district part Step 6 Pressing the RANGE key to select the voltage range Tips z For absolute pressure module the calibrator saves the zero off value and reuses the value automatically Therefore user does not need to zero off the calibration value for each use 9 3 Calibrate 2 Wire Pressure...

Страница 43: ...5 percentage of maxim value of the range the LCD shows ERR in the lower part Pressing the ZERO key initializes the calibrator to 0 and symbol shows on the left upper part of the LCD Figure 14 calibrate 2 wire pressure transmitter Step 5 Apply pressure on the pipe with the pressure source until the desired pressure value displayed on the LCD The LCD shows the present current value from the transmit...

Страница 44: ...atically If it fails to connect the LCD shows NO OP in the upper part Step 3 Press the key to select switch measuring LCD displays switch symbol on the LCD Step 4 Zero off the reading following the pressure module manual When the reading overtops the 5 percentage of maxim value of the range the LCD shows ERR in the lower part Pressing the ZERO key initializes the calibrator to 0 and symbol shows o...

Страница 45: ...y immediately to enter the default set LCD displays the default function on the main districts part and the default value on the assistance district part 10 1 Setting Auto power off time Step 1 Enter the setting state LCD displays AP OFF symbol on the upper part indicating automatic power off setting mode Step 2 Set the time within 0 60 minute range by using the second pair of S T counting from ri...

Страница 46: ...nd Each pair of S T keys corresponds to each digit of the LCD reading Each press of the S T key increases or decreases the digit Increasing the digit from 9 or decreasing it from 0 causes the digit to overflow or underflow allowing you to set the output value without interruption Holding down the S T key continuously changes the digit And the value won t change if it is increased or decreased to t...

Страница 47: ...etting mode Step 2 Shifting between the 50Hz and 60Hz by using the right pair of S T Step 3 Pressing the ON key LCD displays SAVE symbol on the assistance district part for 1s 10 4 Factory default Step 1 Pressing the MODE key LCD displays FACRY symbol on the upper part indicating factory default Step 2 Pressing the ON key LCD displays SAVE symbol on the upper part for 1s All settings are defaulted...

Страница 48: ... Requirements Ambient temperature 23 5 Relative humidity 35 to 75 RH Warm up z Before using warm up the calibrator for the period of time specified z Put the meter into the standard environment for 24 hours and then turn on the power Change the set into non automatic power off state and warm it up for one hour Caution Power Supply new alkaline size AAA battery type 7 is the best choice for adjustm...

Страница 49: ... 3V square wave Applying reference input signals from the calibration standard as listed in the above table Tips z You can also select only the range in need of readjustment to adjust it separately Turn on the meter press the MODE key while simultaneously holding down the ON key into the source adjusting state and then press the MODE key into the measurement adjusting state LCD shows CAL symbol on...

Страница 50: ...ction Step 3 Connect the lead cables to the output terminals of the standard source as shown in Figure16 Step 4 Pressing the RANGE key selects the range The measured value and unit shall be shown in the upper part Figure 16 Calibrating DC voltage and frequency of the LCD Step 5 Pressing the ZERO key enters the measurement CAL mode The LCD shows the present adjusting point CAL 0 in the lower part a...

Страница 51: ... select frequency function Step 3 Connect the lead cables to the output terminals of the standard source as shown in Figure 16 Step 4 The measured value and unit shall be shown in the upper part of the LCD Step 5 Pressing the ZERO key enters the measurement CAL mode The LCD shows the present adjusting point CAL 0 in the lower part and the reference frequency and unit needed for the point in the up...

Страница 52: ... in the upper part of the LCD Step 5 Pressing the ZERO key enters the 50mA of DCmA CAL mode The LCD shows the present adjusting point CAL 0 in the lower part and the reference current and unit needed for the point in the upper part Step 6 Pressing the key saves the adjusted value and the LCD shows SAVE symbol in the upper part for 2 seconds Step 7 Pressing the ZERO key exits the CAL mode The adjus...

Страница 53: ...he set into non automatic power off state and warm it up for one hour Power Supply new alkaline size AAA battery type 7 is the best choice for adjustment Source Adjustment Operation Table 5 Adjustment Points of Source Functions Range Adjustment Point Remarks 0 A F DCV_1000mV 0 0 1000mV DCV_10V 0 10V DCmA_20mA 0 20mA Adjusting the displayed value same with the reading of the digit meter when the pr...

Страница 54: ...elect DC voltage function Connect the lead cables for measurement to the standard digital meter as shown in Figure 18 Step 2 Pressing the RANGE key selects the right range Figure 18 Adjusting voltage source Step 3 The LCD shows CA 0 or CA F symbol on the associate districts part and the calibrator is ready for the zero point or F point adjustment of source functions The LCD shows the highest five ...

Страница 55: ...t digit the digit in the right of the assistance districts LCD part Step 5 Press the 25 100 key to save the CAL adjustment reading Step 6 Pressing the START key shifts to the next setpoint Step 7 By repeating steps 3 to 6 you can adjust all the adjustment point assigned to that range Step 8 By repeating steps 2 to 7 you can adjust all ranges of the DC voltage source function Note z Make sure the p...

Страница 56: ...digits and its unit in the main districts part and the lowest digit of the calibrated sourced value in the right of the assistance districts part respectively Step 3 Read the calibrator output on the calibration standard Then using the pair of S T keys adjust the reading so that it matches the measured CAL adjustment setpoint In the CAL mode the right pair of S T keys are used to increase or decre...

Страница 57: ... Batteries or fuse Warning To avoid possible electric shock remove the test leads from the calibrator before open the battery door And make sure the battery door is tightly closed before turning on the calibrator Caution z To avoid possible linkage of the liquid and explosion of the battery make sure to place the battery with right polarity z Do not operate the battery in short circuit z Do not di...

Страница 58: ...own in Figure 20 With a standard blade hand screwdriver turn each battery door screw a quarter counterclockwise to remove the battery door z Step 3 Replace with four new AAA alkaline batteries under the instructions shown on the Figure 20 Replacing batteries and fuses battery door Or replace the blown fuses with same type F1 50mA 250V F2 50mA 250V or F3 63mA 250V Step 4 Reinstall and tighten the b...

Страница 59: ... battery Step 1 Turn off the calibrator Step 2 Connect the plug into the charging jack of the terminal Step 3 Plug the charger into the AC power Note In normal charging function the indicating light of the charger lights on When finished the charging function stops automatically and the indicating light becomes dark Blinking of the indicating light means the charger is not connected or no battery ...

Страница 60: ...st and replace them if necessary If you suspect that the meter has failed review this manual to make sure you are operating it correctly If the meter still fails to operate properly pack it securely in its original container if available and forward it postage paid to the nearest Service Center The company assumes NO responsibility for damage in transit The Company guarantees a rapid repair and ma...

Страница 61: ...61 15 3 Replacement of Parts All the types of parts are listed in Table 6 see Figure 21 as reference Figure 21 Replacing part ...

Страница 62: ...Spring C 3 5 Terminal Gasket 8 19 AAA Alkaline battery 4 6 LCD Frame 1 20 Tilt stand 1 7 LCD 1 21 Screw M3 16 4 8 Backlight Panel 1 22 Battery Door 1 9 Conductive Rubber wire 2 23 Plastic Screw 2 10 LCD Circuit Panel 1 24 Sponge length width height 40 6 6 1 11 Screw M3 8 4 25 Sponge length width height 48 10 2 5 1 12 Terminal Cover 1 26 Outer Protector 1 13 Cover Door 1 14 Main Circuit Panel 1 ...

Страница 63: ...ressure module and its type see Table 8 and Table 9 For information about the new pressure module which isn t listed in Table 8 and table 9 please contact the representative of the company Table 7 Options No Name of the Options Mode 1 CALCP Pressure module convertor A000018 2 Test Hoop TP907110 3 CA Battery Parcel A000021 4 CA Charger A000020 F Figure 22 Options ...

Страница 64: ...ays OL overload if the reading exceeds those range listed in the following table Table 8 APM S Pressure Module Specification Part number Range Pressure Type 1 Reference Accuracy 2 Entire Accuracy 3APM010WGSG 0to2 49kPa 0to 10 inH20 gauge 0 20 0 30 APM007KGSG 0to7kPa gauge 0 10 0 20 APM035KGSG 0to35kPa gauge 0 035 0 07 APM070KGSG 0to70kPa gauge 0 035 0 07 APM001BGSG 0to100kPa gauge 0 035 0 07 APM16...

Страница 65: ...SG 0to400kPa gauge 0 025 0 05 APM006BGSG 0to600kPa gauge 0 025 0 05 APM010BGSG 0to1MPa gauge 0 025 0 05 APM016BGSG 0to1 6MPa gauge 0 025 0 05 APM021BGSG 0to2 1MPa gauge 0 025 0 05 APM025BGSG 0to2 5MPa gauge 0 025 0 05 APM040BGSG 0to4MPa gauge 0 025 0 05 APM060BGSG 0to6MPa gauge 0 025 0 05 ...

Страница 66: ...0 05 APM250BGSG 0to25MPa gauge 0 025 0 05 APM400BGSG 0to40MPa gauge 0 025 0 05 APM600BGSG 0to60MPa gauge 0 025 0 05 APM700BGSG 0to70MPa gauge 0 025 0 05 APM005PDSG 0to34kPa 0to 5psi Differential 0 035 0 07 APM100PDSG 0to689kPa 0to 100psi Differential 0 025 0 05 APM005PASG 0to34kPa 0to 5psi absolute 0 035 0 07 ...

Страница 67: ...35 0 07 1 Reference Accuracy is defined as the full scale range accuracy gained in the lab environment 2 Entire Accuracy is defined as the full scale range accuracy in one year including 0 50 temperature compensation 3 APM010WGSG pressure module terminal is non isolated which is compatible only with dry or non corrosive air Other pressure module terminal is 316LSS isolation which can be combined w...

Страница 68: ...PM01KPGHG 6890kPa 1000psi gauge APM015PAHG 103kPa 15psi absolute APM050PAHG 345kPa 50psi absolute APM100PAHG 689kPa 100psi absolute APM500PAHG 3450kPa 500psi absolute APM01KPAHG 6890kPa 1000psi absolute APM03KPAHG 20670kPa 3000psi absolute 1 Six month full scale range accuracy includes 15 45 temperature compensation 2 All pressure terminals are non isolated which are compatible only with dry or no...

Страница 69: ...olution Accuracy Remark DCV 200mV 20 00mV 220 00mV 10μV 0 02 0 02 Input Resistance 100MΩ 5V 0 5000V 5 5000V 0 1mV 0 02 0 01 Input Resistance 1MΩ 50V 5 000V 55 000V 1mV 0 03 0 01 DCmA 50mA 4 000mA 55 000mA 1μA 0 02 0 01 Input Resistance 5Ω FREQ 500Hz 3Hz 500 00Hz 0 01Hz 2digit Input Impedance 100 kΩ at least Sensitivity 3Vp p minimum Duty Cycle 50 5KHz 3Hz 5 0000KHz 0 1Hz 50KHz 3Hz 50 000KHz 1Hz CO...

Страница 70: ... applicable accuracy specification per degree ºC for 5 ºC to 18ºC and 28ºC to 50ºC z Maximum voltage between VΩHz terminal and COM terminal 60 Vpk z Maximum Input current 55mA z Protected with a 63mA 250V fast blow fuse General Specifications for Source These specifications assume A 1 year calibration cycle An operating temperature of 18ºC to 28ºC 64 4ºC 82 4ºC Relative humidity of 35 to 70 non_co...

Страница 71: ...2 0 01 Maximum output current 5mA DCA 20mA 0 000mA 22 000mA 1μA 0 02 0 02 External supply for simulate mA 5V 28V Maximum load 1KΩ at 20mA FREQ 100Hz 1 00Hz 110 00Hz 0 01Hz 2 count Output voltage 1 11 Vp p zero base waveform Amplitude accuracy 10 0 5V Maximum load 100 KΩ Duty Cycle 50 1KHz 0 100kHz 1 100KHz 1Hz 10KHz 1 0KHz 11 0KHz 0 1KHz PULSE 100Hz 1 100000 cycles 1cyc _______ Output voltage 1 11...

Страница 72: ...V 10 Maximum current 25 mA Short circuit protected Other feature z Temperature Coefficient 0 1 times the applicable accuracy specification per degree for 5 to 18 and 28 to 50 z Maximum voltage between any output terminal and earth 30Vpk Maximum output current Approximately 25mA ...

Страница 73: ...le Determined by pressure module For more detail refer the pressure module about APM Other feature z Temperature Coefficient 0 1 times the applicable accuracy specification per degree for 5ºC to 18ºC and 28ºC to 50ºC z Maximum voltage between any output terminal and earth 60Vpk Maximum output current Approximately 55mA z Protected with a 63mA 250V fast blow fuse ...

Страница 74: ...regarded as correct Whenever any user finds its mistakes omission etc he or she is requested to contact the manufacturer z The present manufacturer is not liable for any accident and hazard arising from the customer misuse or inadvertent operation z The functions described in this operation instruction should not be used as grounds to apply this product to a particular purpose ...