15
15
1 .2 Safety Precautions
Chapter 1
CAUTION
(9) EMC Statement
Guidance and manufacture’s declaration - electromagnetic immunity
The unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the unit should assure that it is used in such an environment .
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vms
150 kHz to 80 MHz
3 V/m
80 MHz to 2 .5 GHz
[3] V
[3] V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the unit, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter .
Recommended separation distance
d=[1.2]√P
d=[1.2]√P 80 MHz to 800 MHz
d=[2.3]√P 800 MHz to 2.5 GHz
where P is the maximum output power rating
of the S transmitter in watts (W) according to
the transmitter manufacturer and d is the rec-
ommended separation distance in meters (m) .
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range .
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
[NOTE] At 80 MHz and 800 MHz, the higher frequency range applies .
[NOTE] These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mo
-
bile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the unit is used execeeds the appli
-
cable RF compliance level above, the unit should be observed to verify normal operation . If abnormal performance
is observed, additional measures may necessary, such as reorienting or relocation the unit .
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than [3] V/m.
Содержание REGIUS SIGMA2
Страница 1: ...EN 05 0197 Operation Manual DIRECT DIGITIZER REGIUS SIGMA2...
Страница 2: ......
Страница 5: ...5 Introduction...
Страница 10: ...10...
Страница 21: ...21 Chapter 2 Product Outline An explanation of the device s features and functions...
Страница 27: ...27 Chapter 3 Reader Operation An explanation on how to operate this device...
Страница 35: ...35 Chapter 4 Cassette Plate Operation An explanation on cassette and plate handling...
Страница 45: ...45 Chapter 5 Troubleshooting An explanation on how to deal with errors when they occur...
Страница 62: ...62...
Страница 63: ...63 Chapter 6 Maintenance An explanation of maintenance items that must be followed for safe device operation...
Страница 70: ...70...
Страница 71: ...71 Appendix A list of device specifications etc...
Страница 74: ...74...
Страница 75: ......
Страница 76: ...A5ERBA01EN05 2015 08 31 JD...