88
3
En
vir
onment Set
tings/Other Set
tings
User Type
Settings can be protected for each user.
Memo
The user type is set to “Administrator” under the initial settings.
Operating Procedure
Start the procedure from the <Setting> - <Instrument setup> screen.
1
Use [
] or [
] to move the cursor to “User type”,
and then press the [Confirmation] key.
The <User type> setting screen
is displayed.
2
Use [
] or [
] to move the cursor and select the
user type.
Selections
{
Administrator : All settings can be changed.
{
Worker
: Some settings cannot be changed.
Memo
The operations that workers can perform are as follows
• “Sample” browsing/measurement/printing/deletion
• “Target” browsing/printing
• Calibration
• Instrument info. displaying
• Change user type
3
Press the [Confirmation] key to confirm, and then
press [ESC].
The selection is confirmed and the screen returns to the
<Instrument setup> screen
.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Instrument setup> screen.
Sample
Setting
0002 TARGET-2
Instrument setup
User type
Administrator
Language
English
Date format
yyyy/mm/dd
Date & time
2016/08/25
Sample
Setting
0002 TARGET-2
Instrument setup
User type
Administrator
Language
English
Date format
yyyy/mm/dd
Date & time
2016/08/25
User type
Administrator
Worker
Содержание CM-M6
Страница 1: ...Instruction Manual Please read before using the instrument...
Страница 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Страница 102: ...99...
Страница 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Страница 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Страница 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Страница 126: ...123...
Страница 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...