32
2
Measur
ement
Calibration
The following two types of calibration can be performed with this instrument.
•
White calibration : The reflectance is measured in advance using a known white calibration plate in order to
provide a reflectance scale.
•
Zero calibration : Only the amount of stray light is measured with the zero calibration box in advance in order
to eliminate the effects of the stray light.
Zero Calibration
Since this instrument stores the data of the previous zero calibration, repeating zero calibration every time the
instrument is turned ON is not necessary. If, however, the temperature or other measurement conditions change
greatly, or if the instrument is not used for a long period of time, zero calibration must be performed before white
calibration.
Memo
•
The effects of stray light (i.e., light generated due to the flare characteristics of the optical system) will be
compensated for automatically by the zero calibration data.
•
The amount of stray light may change because of dust or dirt that has collected in the optical system, humidity,
repeated operation, or vibration and shock exerted on the instrument. In this case, performing zero calibration
periodically is recommended.
Notes
• If the instrument is not used for long periods of time, the zero calibration data stored in the instrument may
be lost. If the data is lost, zero calibration must be performed again.
Operating Procedure
Start the procedure from the measurement screen.
1
Press [MENU], and then use [
] or [
] to display the
<Calibration> screen.
Memo
The following screen prompting calibration is displayed
when the instrument is turned ON. To perform zero
calibration, select “Calibration (Including Zero Cal.)”. After
completing zero calibration, perform white calibration.
Calibration
White calibration
Zero calibration date
2016/08/22 13:51:58
White calibration date
2016/08/23 17:14:06
White ID
1000001
Calibration
(Including Zero Cal.)
When power is turned ON
2
Use [
] or [
] to move the cursor to “Calibration
(Including Zero Cal.)”, and then press the
[Confirmation] key.
Sample
Calibration
White calibration
White ID
1000001
White calibration date
2016/08/23 17:14:06
Zero calibration date
2016/08/22 13:51:58
(Including Zero Cal.)
Calibration
Содержание CM-M6
Страница 1: ...Instruction Manual Please read before using the instrument...
Страница 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Страница 102: ...99...
Страница 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Страница 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Страница 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Страница 126: ...123...
Страница 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...