Auto Average (1 to 10)
Specify the number of measurements for auto averaging. Every time the measurement key/button is pressed, the
average of the data obtained from the specified number of continuous measurements is determined as sample
data.
Memo
“1 Times” is set as the number of measurements for automatic averaging under the initial settings.
Operating Procedure
Start the procedure from the <Measurement cond.> - <Measurement setup> screen.
1
Use [
] or [
] to move the cursor to “Auto
average”, and then press the [Confirmation] key.
2
and
will be displayed above and below the
count in the count selection. Use [
] or [
] to
specify a value.
Settings
{
01 to 10 Times:
Specify the number of measurements, from 1 to 10, to
perform when using auto averaging.
3
Press the [Confirmation] key.
The selection is confirmed and the screen returns to the
<Measurement setup> screen
.
Notes
If [ESC] is pressed without pressing the [Confirmation]
key, the settings will not be changed and the screen will
return to the <Measurement setup> screen.
Sample
Measurement cond.
Measurement setup
Measurement angle
Manual average option
Manual save
Manual average
1 Times
Auto average
1 Times
Sample
Measurement cond.
Measurement setup
Measurement angle
Manual average option
Manual save
Manual average
Auto average
Auto average
0
3
/ 1 0
Содержание CM-M6
Страница 1: ...Instruction Manual Please read before using the instrument...
Страница 18: ...15 Names and Functions of Parts 1 8 9 4 6 5 2 3 10 11 12 14 13 7...
Страница 102: ...99...
Страница 119: ...116 5 Troubleshooting Chapter 5 Troubleshooting Message List 117 Troubleshooting 119...
Страница 123: ...120 6 Appendix Chapter 6 Appendix Main Specifications 121 Dimensions 122...
Страница 125: ...122 6 Appendix Dimensions Unit mm 50 M6 1 0 DEPTH 6 M6 1 0 DEPTH 6 55 33 152 3 49 5 81 239...
Страница 126: ...123...
Страница 128: ...En 9222 A9DT 13 2016 KONICA MINOLTA INC BIFCGK...