KV5000 Instrument & Software 

Operation and Instruction Manual 

 

KV5000-Manual

    

 

-

21-

 

6  Safety Features

 

 

The  Koehler  KV5000  Kinematic  Viscosity 

System  is  equipped  with  several  safety  and 
protection  features,  which  are  described  in  the 
following sections. 

 

WARNING:

  The  bath  is 

NOT

  Explosion 

Proof.  

 

6.1  Over-Temperature Protection

 

 

The  Koehler  KV5000  Kinematic  Viscosity 

System  is  equipped  with  Over-temperature 
Protection (OTP) circuitry,  which prevents the unit 
from  exceeding  unsafe  operating  temperatures.  If 
the unit cannot maintain the set point temperature 
and begins to decline, the OTP circuitry may have 
been activated. Please follow these steps. 

 

1.  Turn off the unit  power switch and disconnect 

the line cord. 
 

2.  Determine  the  source  of  the  problem  and 

correct the situation. 
 

3.  Restart  the  unit.  Monitor  the  operations  to 

ensure  that  the  unit  is  operating  properly.  If 
you  are  still  experiencing  trouble,  please 
contact 

Koehler 

technical 

service 

for 

assistance. 

 

6.2  Over-Power Protection

 

 

The  Koehler  KV5000  Kinematic  Viscosity  System 
is  equipped  with  Over-power  Protection  circuitry, 
which  prevents  the  unit  from  unsafe  electrical 
conditions. If power to the unit is lost, then turn off 
the  main  power  and  turn  it  back  on  again.  The 
main  power  switch  also  functions  as  a  circuit 
breaker. 

 

7  Troubleshooting

 

 

WARNING:

 

Troubleshooting  procedures  involve 

working  with  high  voltages  and/or  temperatures 
which  may  result  in  personal  injury  or  death,  and 
should  only  be  performed  by  trained  personnel. 
Please  do  not  hesitate  to  contact  Koehler  for 
assistance. 
 

 

7.1  The unit does not power up

 

 

 

Establish  that  the  socket  outlet  is  providing 
proper and adequate voltage. 

 

Check  if  Overpower  Protection  circuitry 
located  directly  behind  the  temperature 
controller  inside  the  front  tray  has  been 
activated. 

 

Check if line switch is in the 

ON 

position. 

 

If  problem  persists,  please  call  the  Koehler 
technical service department for assistance. 

 

7.2  The unit is an and keeps resetting 

into start up routine

 

 

 

For  220V  units,  make  sure  that  the  socket 
outlet is greater than 215V. 

 

Check  if  there  is  a  steady  and  reliable  power 
source. 

 

Make  sure  the  connector  plug  on  the  rear 
panel is firmly plugged in. 

 

7.3  The unit is on but the bath does not 

heat up

 

 

 

Make sure that the actual temperature reading 
is not higher than the set-point temperature. 

 

Determine  if  the  temperature  controller  is 
properly calibrated by comparison to an ASTM 
standard thermometer. 

 

Determine  if  the  Overtemperature  Protection 
(OTP) circuitry as been activated. 

 

7.4  The bath heats up but does not 

stabilize

 

 

 

Make  sure  there  are  no  drafts  from  open 
doors,  windows,  or  environmental  control 
vents in the vicinity of the bath. 

 

If the set temperature is close to ambient, then 
you  may  need  to  circulate  cold  water  through 
the  coils  in  order  to  achieve  proper 
temperature stability

 

8  Maintenance

 

 

WARNING.

 

Disconnect  power  to  the  unit  before 

servicing  to  avoid  exposure  to  high  voltages 
and/or  temperatures  which may  result  in  personal 
injury  or  death.  If  you  have  any  questions  about 
maintaining  your  equipment,  then  please  do  not 
hesitate  to  contact  the  Koehler  technical  service 
department. 

 

8.1  Routine Maintenance

 

 

The  KV5000  Kinematic  Viscosity  System 

requires little routine maintenance to provide many 
years  of  continuous  service.  However,  over  the 
course  of  time,  some  instrument  parts  may  need 
to  be  replaced.  When  ordering  replacement 
part(s),  please  provide  the  model  number,  serial 

Содержание KV5000

Страница 1: ...ia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrument...

Страница 2: ......

Страница 3: ...h the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it leav...

Страница 4: ......

Страница 5: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 1...

Страница 6: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 2...

Страница 7: ...is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk e...

Страница 8: ...mum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent...

Страница 9: ...tion 14 5 1 Bath 14 5 2 Power 15 5 3 Engage the Viscometer in Optical Sensor Assembly 15 5 4 Adjusting the Sensors 17 5 5 Control Panel Operation 18 5 6 Test Procedure 19 6 Safety Features 21 6 1 Over...

Страница 10: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 6 10 Service 27 11 Storage 27 12 Warranty 27 13 Returned Goods Policy 27 Notes 28...

Страница 11: ...range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or tec...

Страница 12: ...s Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certification...

Страница 13: ...box and remove packing foam 6 Extract instrument and place on suitable cart for transportation to work area lab bench WARNING Be sure two or more individuals are available for extracting and lifting...

Страница 14: ...quipment WARNING Certain mechanical stresses applied to the tempered glass front panel and inner the inner borosilicate glass jar pose a hazard to personnel Impacts to these during normal operation ca...

Страница 15: ...t This port allows for independent temperature measurement of the bath temperature with a thermometer or a Pt 100 RTD probe for precise temperature measurements and digital temperature controller cali...

Страница 16: ...8 500 C02 OS 500 8 1 600 to 8 000 378 600 C02 OS 600 20 4 000 to 20 000 378 650 C02 OS 650 45 9 000 to 45 000 378 700 C02 OS 700 100 20 000 to 100 000 Ubbelohde Viscometers Ubbelohde viscometers measu...

Страница 17: ...eter holders for use with the KV 5000 The correct holder must be used with the corresponding viscometer tube for proper operation Viscometer Tube Type Corresponding Holder Cannon Fenske Routine K23381...

Страница 18: ...ol to learn the heating parameters needed for any particular set point temperature This operation should be done when installing a new unit after replacing or changing the bath medium type or utilizin...

Страница 19: ...or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed Se...

Страница 20: ...e Figure 8 Viscometer Tube Completely Inserted in the Optical Sensor Assembly Optical wires are not shown for illustrative purposes only If the optical sensor assembly has a Teflon retaining ring rath...

Страница 21: ...d the receiver When the signal can pass between the two sensors the upper orange light will be on When the signal is interrupted when a meniscus passes through the orange light will flicker or turn of...

Страница 22: ...he viscometer parameters displayed 4 Test Data Displays the run time during a test in seconds Time 1 indicates the elapsed time for a standard test for a viscometer with a single constant The Time 2 i...

Страница 23: ...test procedure listed in section 1 2 of this document when performing Kinematic viscosity testing After the main power switch to the instrument is turned on the control panel will illuminate and disp...

Страница 24: ...n The status will automatically change from RUN to RUN 2 Once the sample passes the second indicator line on the viscometer tube the TIME 2 will automatically stop the status will change from RUN 2 to...

Страница 25: ...s providing proper and adequate voltage Check if Overpower Protection circuitry located directly behind the temperature controller inside the front tray has been activated Check if line switch is in t...

Страница 26: ...ose of this material and its container to a hazardous of special waste collection point Country State or local laws may impose additional regulatory requirements regarding disposal Dispose in accordan...

Страница 27: ...from the main instrument it is now easy to access the inside of the bath jar for further cleaning 16 With the top assembly resting on its back it is now possible to clean the heating coils stirrer bl...

Страница 28: ...low Blow Fuse 0 25A 5x20 mm 278 003 001 Slow Blow Fuse 3 15A 5x20mm 278 020 004 Time Delay Fuse 20A 600VAC 278 102 003 Fast Acting Fuse 0 5A 2AG 275 103 044 Temperature Controller 100 240V with RS 232...

Страница 29: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 25 9 Wiring Diagrams 9 1 Unit Wiring...

Страница 30: ...t for Amp 3 3 Brown 9 9 Blue Y3 V White 10 10 Blue Black X6 Output for Amp1 White 11 11 Blue White X7 Output for Amp 2 White 12 12 Blue Red X10 Output for Amp 3 4 Brown 13 13 White Y4 V White 14 14 Wh...

Страница 31: ...f the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replac...