Make  sure  there  is  sufficient  space  with 
cleaning  clothes  to  place  the  top  assembly 
when lifted off the bath.  

 

17. 

Lift  the  top  assembly  with  one  hand  on  the 
back  side  and  one  hand  on  the  front  side  of 
the top assembly. Place top plate assembly on 
flat surface with the back side of the top plate 
assembly and the heating coils on the surface.  

 

18. 

With  the  top  plate  assembly  now  removed 
from  the  main  instrument  it  is  now  easy  to 
access  the  bath  jar  to  remove  the  protective 
foam insert.  

 

19. 

Re-assemble the instrument 

 

20. 

Ensure  that  all  parts  listed  on  the  packing  list 
are  present.  Inspect  the  unit  and  all 
accessories  for  damage.  If  any  damage  is 
found,  keep  all  packing  materials  and 
immediately  report  the  damage  to  the  carrier. 
We  will  assist  you  with  your  claim,  if 
requested.  When  submitting  a  claim  for 
shipping  damage,  request  that  the  carrier 
inspect the shipping container and equipment. 
Do not return goods to Koehler without written 
authorization.  

 

3.3  Setup 

 

Equipment Placement.

 Place the instrument on a 

firm,  level  table  in  an  area  with  adequate 
ventilation  or  in  a  hood.  The  unit  may  be  leveled 
by  making  minor  turning  adjustments  to  the  feet 
located  at  the  base  of  the  unit.  Please  note  that 
Koehler  does  not  supply  a  level  with  this 
equipment. 
 

WARNING:

 

Certain  mechanical  stresses 

applied  to  the  tempered  glass  front  panel  and 
inner the inner borosilicate glass jar pose a hazard 
to  personnel.  Impacts  to  these  during  normal 
operation  can  shatter  the  glass  components 
releasing  the  bath  fluid  posing  thermal  burn 
hazards and slipping hazards. For this reason the 
instrument  must  be  operated  by  and  accessible 
only  to  trained  and  authorized  personnel  and 
located  in  an  area  where  this  type  of  impact  is 
likely to occur.  

 
Environmental  Conditions: 

The  instrument 

environment  must  comply  with  the  following 
conditions for proper setup: 

 

  No / Low Dust 

 

  No direct sunlight 

 

  Not near heating or AC ventilation ducts 

 

  No Vibrations 

 

  Clearance from other instruments 

 

  Temperature Range: 5 to 40°C 

 

  Elevation to 2000 meters 

 

Relative Humidity: < 80% 
 

Ventilation.

  A  fume  hood  or  exhaust  system  is 

required  when  operating  the  unit.  Flammable 
vapors  and/or  steam  are  generated  during 
operation  and  must  not  be  permitted  to 
accumulate.  A  canopy-style  hood  may  be  used  if 
the height from the top of the unit to the canopy is 
5  feet  or  less.  The  exhaust  blower  should  have  a 
rating of 1000 C.F.M. or greater. 
 

Power. 

Connect  the  line  cords  to  properly  fused 

and grounded receptacles with the correct voltage 
as  indicated  in  section  1.3  or  on  the  back  of  the 
unit.  
 

WARNING:

   

For  safety,  disconnect  the 

power  when  performing  any  maintenance  and/or 
cleaning.

 

Do 

NOT

  turn  the  power  on  unless  the 

bath  is  filled  with  the  proper  medium;  otherwise, 
damage  may  occur  to  the  unit  and  the  warranty 
will be void. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание KV5000

Страница 1: ...ia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrument...

Страница 2: ......

Страница 3: ...h the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it leav...

Страница 4: ......

Страница 5: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 1...

Страница 6: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 2...

Страница 7: ...is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk e...

Страница 8: ...mum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent...

Страница 9: ...tion 14 5 1 Bath 14 5 2 Power 15 5 3 Engage the Viscometer in Optical Sensor Assembly 15 5 4 Adjusting the Sensors 17 5 5 Control Panel Operation 18 5 6 Test Procedure 19 6 Safety Features 21 6 1 Over...

Страница 10: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 6 10 Service 27 11 Storage 27 12 Warranty 27 13 Returned Goods Policy 27 Notes 28...

Страница 11: ...range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or tec...

Страница 12: ...s Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certification...

Страница 13: ...box and remove packing foam 6 Extract instrument and place on suitable cart for transportation to work area lab bench WARNING Be sure two or more individuals are available for extracting and lifting...

Страница 14: ...quipment WARNING Certain mechanical stresses applied to the tempered glass front panel and inner the inner borosilicate glass jar pose a hazard to personnel Impacts to these during normal operation ca...

Страница 15: ...t This port allows for independent temperature measurement of the bath temperature with a thermometer or a Pt 100 RTD probe for precise temperature measurements and digital temperature controller cali...

Страница 16: ...8 500 C02 OS 500 8 1 600 to 8 000 378 600 C02 OS 600 20 4 000 to 20 000 378 650 C02 OS 650 45 9 000 to 45 000 378 700 C02 OS 700 100 20 000 to 100 000 Ubbelohde Viscometers Ubbelohde viscometers measu...

Страница 17: ...eter holders for use with the KV 5000 The correct holder must be used with the corresponding viscometer tube for proper operation Viscometer Tube Type Corresponding Holder Cannon Fenske Routine K23381...

Страница 18: ...ol to learn the heating parameters needed for any particular set point temperature This operation should be done when installing a new unit after replacing or changing the bath medium type or utilizin...

Страница 19: ...or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed Se...

Страница 20: ...e Figure 8 Viscometer Tube Completely Inserted in the Optical Sensor Assembly Optical wires are not shown for illustrative purposes only If the optical sensor assembly has a Teflon retaining ring rath...

Страница 21: ...d the receiver When the signal can pass between the two sensors the upper orange light will be on When the signal is interrupted when a meniscus passes through the orange light will flicker or turn of...

Страница 22: ...he viscometer parameters displayed 4 Test Data Displays the run time during a test in seconds Time 1 indicates the elapsed time for a standard test for a viscometer with a single constant The Time 2 i...

Страница 23: ...test procedure listed in section 1 2 of this document when performing Kinematic viscosity testing After the main power switch to the instrument is turned on the control panel will illuminate and disp...

Страница 24: ...n The status will automatically change from RUN to RUN 2 Once the sample passes the second indicator line on the viscometer tube the TIME 2 will automatically stop the status will change from RUN 2 to...

Страница 25: ...s providing proper and adequate voltage Check if Overpower Protection circuitry located directly behind the temperature controller inside the front tray has been activated Check if line switch is in t...

Страница 26: ...ose of this material and its container to a hazardous of special waste collection point Country State or local laws may impose additional regulatory requirements regarding disposal Dispose in accordan...

Страница 27: ...from the main instrument it is now easy to access the inside of the bath jar for further cleaning 16 With the top assembly resting on its back it is now possible to clean the heating coils stirrer bl...

Страница 28: ...low Blow Fuse 0 25A 5x20 mm 278 003 001 Slow Blow Fuse 3 15A 5x20mm 278 020 004 Time Delay Fuse 20A 600VAC 278 102 003 Fast Acting Fuse 0 5A 2AG 275 103 044 Temperature Controller 100 240V with RS 232...

Страница 29: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 25 9 Wiring Diagrams 9 1 Unit Wiring...

Страница 30: ...t for Amp 3 3 Brown 9 9 Blue Y3 V White 10 10 Blue Black X6 Output for Amp1 White 11 11 Blue White X7 Output for Amp 2 White 12 12 Blue Red X10 Output for Amp 3 4 Brown 13 13 White Y4 V White 14 14 Wh...

Страница 31: ...f the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replac...

Страница 32: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 28 Notes...

Страница 33: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 29 Notes...

Страница 34: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 30 Notes...

Отзывы: