KV5000 Instrument & Software 

Operation and Instruction Manual 

 

KV5000-Manual

    

 

-

14-

 

Setting  the  Temperature

.  Set  the  desired 

operating  temperature  by  adjusting  the  set 
point with the up and down keys. The set point 
will  be  displayed  in  the  lower  green  Set  Point 
LED display and the actual temperature will be 
displayed  in  the  upper  red  Process  LED 
display.  Please  allow  the  instrument  to  fully 
equilibrate before proceeding with any testing. 

 

Temperature Calibration.

  This routine allows the 

digital  temperature  controller  to  be  calibrated 
to  a  certified  thermometer.  In  the  calibration 
mode,  the  display  will  always  give  2  decimal 
places. 

 

a.  Use  a  certified  calibrated  measuring 

device  to 

acquire  the 

temperature. 

Calculate  the  difference  between  the 
measuring  device  and  the  Process  value 
displayed on the controller. 

b.  Press the program key two times until 

PCt

 

is  displayed  in  the  lower  green  LED 
display.  Press  the  DOWN  key.  CAL  will 
display on the lower green display. If there 
is  a  value  observed  in  the  upper  red  LED 
display, add  it  to the calculated difference 
obtained  in  the  previous  step.  This  is  the 
offset value. 

 

c.  Press  the  Program  Key.  The  lower  green 

display will flash. Use the up or down keys 
to  adjust  to  the  new  calibration  offset 
value  on  the  upper  red  display  calculated 
in  the  previous  step.  When  the  value  has 
been 

entered, 

the 

controller 

will 

automatically  store  the  value.  The  lower 
green  display  will  stop  flashing.  If  further 
adjustments  are  necessary,  press  the 
Program  Key  again.  Resume  regular 
operations by pressing  the Exit / Function 
key two times. Verify if the new calibration 
is  correct  by  observing  the  upper  red 
display  and  comparing  the  value  with  the 
calibrated reference device.  

 
 

Auto  Tune.

  This  routine  allows  the  digital 

temperature  control  to  learn  the  heating 
parameters needed for any particular set point 
temperature.  This  operation  should  be  done 
when  installing  a  new  unit,  after  replacing  or 
changing  the  bath  medium  type,  or  utilizing  a 
different  temperature  set  point  20%  different 
from 

the 

previously 

used 

set 

point 

temperature. 

 

a.  Set  the  operating  temperature  to  the 

desired setting. 

 
b.  Press  the  up  and  down  arrow  buttons 

simultaneously for about 5 seconds. When 
Auto  Tune  is  active,  the  lower  green  LED 
display  will  blink 

TUNE

.  Auto  Tune  will 

automatically toggle off when the set point 
temperature is reached. Auto tune can be 
terminated  by  pressing  the  up  &  down 
buttons simultaneously again.  

 

4.4  Recommended Accessories 

 

 

Withdrawal Bulb

. Use to pull sample into 

viscometer tube. 

Part number: K22090 

 

 

Rubber Stopper

. Use to plug viscometer 

tube and hold up sample prior to testing.  

Part number: K23311 

 

 

Operation near or below ambient 
temperatures.

 A water supply or external 

chiller may be used to maintain the 
temperature of the bath for cooler 
temperature. Hose-barb connections for 
water or refrigerated coolant are located at 
the top of the bath. There are two 
connections: inlet and outlet. The hose 
connections can be secured with copper 
wire to prevent the tubing from disengaging 
from the connection.

 

 

5  Operation 

 

This  unit  is  designed  for  performing  kinematic 
viscosity  tests  using  glass  capillary  viscometers. 
Please  be  sure  to  read  the  safety  and  hazard 
warnings,  the  installation  procedure,  and  any  of 
the  standard  test  methods  before  operating  this 
instrument. 

 

5.1  Bath 

 

Fill  the

 

bath  with  the  appropriate  heat  transfer 

fluid  based  upon  the  testing  temperature.  Fill  the 
bath with the medium from the top through one of 
the seven port openings using a wide neck funnel 
with a 500mL griffin beaker adding about ½ L at a 
time to 2" (5 cm) from the top of the bath to allow 
room  for  fluid  expansion.  This  will  provide  the 
proper  depth  for  immersing  the  viscometers  and 
allow for thermal expansion. 

 

Содержание KV5000

Страница 1: ...ia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrument...

Страница 2: ......

Страница 3: ...h the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it leav...

Страница 4: ......

Страница 5: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 1...

Страница 6: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 2...

Страница 7: ...is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk e...

Страница 8: ...mum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent...

Страница 9: ...tion 14 5 1 Bath 14 5 2 Power 15 5 3 Engage the Viscometer in Optical Sensor Assembly 15 5 4 Adjusting the Sensors 17 5 5 Control Panel Operation 18 5 6 Test Procedure 19 6 Safety Features 21 6 1 Over...

Страница 10: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 6 10 Service 27 11 Storage 27 12 Warranty 27 13 Returned Goods Policy 27 Notes 28...

Страница 11: ...range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or tec...

Страница 12: ...s Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certification...

Страница 13: ...box and remove packing foam 6 Extract instrument and place on suitable cart for transportation to work area lab bench WARNING Be sure two or more individuals are available for extracting and lifting...

Страница 14: ...quipment WARNING Certain mechanical stresses applied to the tempered glass front panel and inner the inner borosilicate glass jar pose a hazard to personnel Impacts to these during normal operation ca...

Страница 15: ...t This port allows for independent temperature measurement of the bath temperature with a thermometer or a Pt 100 RTD probe for precise temperature measurements and digital temperature controller cali...

Страница 16: ...8 500 C02 OS 500 8 1 600 to 8 000 378 600 C02 OS 600 20 4 000 to 20 000 378 650 C02 OS 650 45 9 000 to 45 000 378 700 C02 OS 700 100 20 000 to 100 000 Ubbelohde Viscometers Ubbelohde viscometers measu...

Страница 17: ...eter holders for use with the KV 5000 The correct holder must be used with the corresponding viscometer tube for proper operation Viscometer Tube Type Corresponding Holder Cannon Fenske Routine K23381...

Страница 18: ...ol to learn the heating parameters needed for any particular set point temperature This operation should be done when installing a new unit after replacing or changing the bath medium type or utilizin...

Страница 19: ...or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed Se...

Страница 20: ...e Figure 8 Viscometer Tube Completely Inserted in the Optical Sensor Assembly Optical wires are not shown for illustrative purposes only If the optical sensor assembly has a Teflon retaining ring rath...

Страница 21: ...d the receiver When the signal can pass between the two sensors the upper orange light will be on When the signal is interrupted when a meniscus passes through the orange light will flicker or turn of...

Страница 22: ...he viscometer parameters displayed 4 Test Data Displays the run time during a test in seconds Time 1 indicates the elapsed time for a standard test for a viscometer with a single constant The Time 2 i...

Страница 23: ...test procedure listed in section 1 2 of this document when performing Kinematic viscosity testing After the main power switch to the instrument is turned on the control panel will illuminate and disp...

Страница 24: ...n The status will automatically change from RUN to RUN 2 Once the sample passes the second indicator line on the viscometer tube the TIME 2 will automatically stop the status will change from RUN 2 to...

Страница 25: ...s providing proper and adequate voltage Check if Overpower Protection circuitry located directly behind the temperature controller inside the front tray has been activated Check if line switch is in t...

Страница 26: ...ose of this material and its container to a hazardous of special waste collection point Country State or local laws may impose additional regulatory requirements regarding disposal Dispose in accordan...

Страница 27: ...from the main instrument it is now easy to access the inside of the bath jar for further cleaning 16 With the top assembly resting on its back it is now possible to clean the heating coils stirrer bl...

Страница 28: ...low Blow Fuse 0 25A 5x20 mm 278 003 001 Slow Blow Fuse 3 15A 5x20mm 278 020 004 Time Delay Fuse 20A 600VAC 278 102 003 Fast Acting Fuse 0 5A 2AG 275 103 044 Temperature Controller 100 240V with RS 232...

Страница 29: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 25 9 Wiring Diagrams 9 1 Unit Wiring...

Страница 30: ...t for Amp 3 3 Brown 9 9 Blue Y3 V White 10 10 Blue Black X6 Output for Amp1 White 11 11 Blue White X7 Output for Amp 2 White 12 12 Blue Red X10 Output for Amp 3 4 Brown 13 13 White Y4 V White 14 14 Wh...

Страница 31: ...f the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replac...

Страница 32: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 28 Notes...

Страница 33: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 29 Notes...

Страница 34: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 30 Notes...

Отзывы: