KV5000 Instrument & Software 

Operation and Instruction Manual 

 

KV5000-Manual

    

 

-

20-

 

Press the Viscometer Data Box (section 5.3, point 
3)  to  access  the  Viscometer  Setup  Screen.  See 
Figure  4  in  section  5.3  and  Figure  6  below  for  an 
additional illustration.  
 

 

 

Figure 4. Control Panel_Viscometer Setup Screen

 

 
Refer  to  Section  5.3,  points  5  thru  12  for 
instructions  on  how  to  input  and  load  viscometer 
parameter data.  
 
After  all  required  parameters  are  entered  and  the 
sample  temperature  has  reached  the  test 
temperature it is time to begin the test.  
 

For Automatic Operation:  

 

From  the  Main  Screen,  press  the 

Auto  Mode

 

button.  
 

Press  the  Start  button. 

.  The  viscometer 

status field located at the upper right hand side of 
the screen will change from 

OFF

 to 

READY

.  

 
Remove the stopper. When the sample passes the 
first  indication  line  of  the  viscometer  tube  the 
instrument  status  will  automatically  change  to 

RUN

 and the integrated timer will begin.  

 
When  the  sample  passes  the  second  indication 
line the time will automatically stop, the status will 
display 

DONE

,  and  a  viscosity  value  will 

automatically  be  displayed  at  the  bottom  of  the 
Test Data area (Figure 3, point 4).   
 

Pressing  and  holding  the  Stop  button 

  for  2 

seconds  or  more  will  reset  the  time  and  viscosity 
values and OFF will display as the status.  
 

When  using  a  viscometer  with  two  (2)  constants, 
two (2) time values are used to calculate viscosity.  
 
When the sample passes the first indicator line the 

Time 1

 will stop and 

Time 2

 will begin. The status 

will automatically change from 

RUN

 to 

RUN 2

.  

 
Once the sample passes the second indicator line 
on  the  viscometer  tube  the 

TIME  2

  will 

automatically  stop,  the  status  will  change  from 

RUN  2

  to 

DONE 

and  the  viscosity  value  will 

automatically be calculated and displayed.  
 

For Manual Operation:  

 

From  the  Main  Screen,  press  the 

Manual  Mode

 

button.  
 
Remove the stopper. When the sample passes the 
first  indication  line  of  the  viscometer  tube,  press 

the Start

 

button 

 on the Control Panel. When 

the  viscometer  is  in  standby  mode  the  status  will 
display 

OFF

  at  the  upper  right  hand  side  of  the 

screen. Upon  pressing  the  Start  button 

  the 

instrument  status  will  change  to 

RUN

  and  the 

integrated timer will begin.  
 
When  the  sample  passes  the  second  indication 

line,  press  the  Stop  button 

.  The  time  will 

stop, the status will display 

DONE

, and a viscosity 

value will automatically be displayed at the bottom 
of the Test Data area (Figure 3, point 4).   
 

Pressing  and  holding  the  Stop  button 

  for  2 

seconds  or  more  will  reset  the  time  and  viscosity 
values and OFF will display as the status.  
 
When  using  a  viscometer  with  two  (2)  constants, 
two (2) time values are used to calculate viscosity. 
This  means  that  when  running  a  viscosity  test 
using  a  two  constant  viscometer  the  Stop  button 

 must be pressed twice.  

 
When the sample passes the first indicator line the 

Time 1

 will stop and 

Time 2

 will begin. The status 

will change from 

RUN

 to 

RUN 2

.  

 
Once the sample passes the second indicator line 
on  the  viscometer  tube,  press  the  Stop  button  

 a second time and the 

TIME 2

 will stop. The 

status  will  change  from 

RUN  2

  to 

DONE 

and  the 

viscosity value will automatically be calculated and 
displayed. 
 

Содержание KV5000

Страница 1: ...ia New York 11716 1796 USA Toll Free 1 800 878 9070 US only Tel 1 631 589 3800 Fax 1 631 589 3815 http www koehlerinstrument com e mail info koehlerinstrument com Petroleum Testing Analysis Instrument...

Страница 2: ......

Страница 3: ...h the applicable standards set forth in this certification Specifications ASTM D445 ASTM D2532 ASTM D6074 ASTM D6158 IP 71 ISO 3104 DIN 51550 FTM 791 305 NF T 60 100 This unit is tested before it leav...

Страница 4: ......

Страница 5: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 1...

Страница 6: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 2...

Страница 7: ...is given cataloging the items that require separation from the equipment along with basic information on subsequent processing or recycling prior to disposal of the equipment http www dpa system dk e...

Страница 8: ...mum the following substances preparations and components have to be removed from any separately collected WEEE Mercury containing components such as switches or backlighting lamps compact fluorescent...

Страница 9: ...tion 14 5 1 Bath 14 5 2 Power 15 5 3 Engage the Viscometer in Optical Sensor Assembly 15 5 4 Adjusting the Sensors 17 5 5 Control Panel Operation 18 5 6 Test Procedure 19 6 Safety Features 21 6 1 Over...

Страница 10: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 6 10 Service 27 11 Storage 27 12 Warranty 27 13 Returned Goods Policy 27 Notes 28...

Страница 11: ...range of accurate and reliable instrumentation as well as support for your laboratory testing programs Please do not hesitate to contact us at any time with your inquiries about equipment tests or tec...

Страница 12: ...s Any modification or alteration of this equipment from that of factory specifications is not recommended voids the manufacturer warranty product safety performance specifications and or certification...

Страница 13: ...box and remove packing foam 6 Extract instrument and place on suitable cart for transportation to work area lab bench WARNING Be sure two or more individuals are available for extracting and lifting...

Страница 14: ...quipment WARNING Certain mechanical stresses applied to the tempered glass front panel and inner the inner borosilicate glass jar pose a hazard to personnel Impacts to these during normal operation ca...

Страница 15: ...t This port allows for independent temperature measurement of the bath temperature with a thermometer or a Pt 100 RTD probe for precise temperature measurements and digital temperature controller cali...

Страница 16: ...8 500 C02 OS 500 8 1 600 to 8 000 378 600 C02 OS 600 20 4 000 to 20 000 378 650 C02 OS 650 45 9 000 to 45 000 378 700 C02 OS 700 100 20 000 to 100 000 Ubbelohde Viscometers Ubbelohde viscometers measu...

Страница 17: ...eter holders for use with the KV 5000 The correct holder must be used with the corresponding viscometer tube for proper operation Viscometer Tube Type Corresponding Holder Cannon Fenske Routine K23381...

Страница 18: ...ol to learn the heating parameters needed for any particular set point temperature This operation should be done when installing a new unit after replacing or changing the bath medium type or utilizin...

Страница 19: ...or volatilize producing flammable vapors below the maximum instrument operating temperature 25 C NOTE Coloration of the Mineral or Silicon Bath Oil is an indication that the Medium must be changed Se...

Страница 20: ...e Figure 8 Viscometer Tube Completely Inserted in the Optical Sensor Assembly Optical wires are not shown for illustrative purposes only If the optical sensor assembly has a Teflon retaining ring rath...

Страница 21: ...d the receiver When the signal can pass between the two sensors the upper orange light will be on When the signal is interrupted when a meniscus passes through the orange light will flicker or turn of...

Страница 22: ...he viscometer parameters displayed 4 Test Data Displays the run time during a test in seconds Time 1 indicates the elapsed time for a standard test for a viscometer with a single constant The Time 2 i...

Страница 23: ...test procedure listed in section 1 2 of this document when performing Kinematic viscosity testing After the main power switch to the instrument is turned on the control panel will illuminate and disp...

Страница 24: ...n The status will automatically change from RUN to RUN 2 Once the sample passes the second indicator line on the viscometer tube the TIME 2 will automatically stop the status will change from RUN 2 to...

Страница 25: ...s providing proper and adequate voltage Check if Overpower Protection circuitry located directly behind the temperature controller inside the front tray has been activated Check if line switch is in t...

Страница 26: ...ose of this material and its container to a hazardous of special waste collection point Country State or local laws may impose additional regulatory requirements regarding disposal Dispose in accordan...

Страница 27: ...from the main instrument it is now easy to access the inside of the bath jar for further cleaning 16 With the top assembly resting on its back it is now possible to clean the heating coils stirrer bl...

Страница 28: ...low Blow Fuse 0 25A 5x20 mm 278 003 001 Slow Blow Fuse 3 15A 5x20mm 278 020 004 Time Delay Fuse 20A 600VAC 278 102 003 Fast Acting Fuse 0 5A 2AG 275 103 044 Temperature Controller 100 240V with RS 232...

Страница 29: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 25 9 Wiring Diagrams 9 1 Unit Wiring...

Страница 30: ...t for Amp 3 3 Brown 9 9 Blue Y3 V White 10 10 Blue Black X6 Output for Amp1 White 11 11 Blue White X7 Output for Amp 2 White 12 12 Blue Red X10 Output for Amp 3 4 Brown 13 13 White Y4 V White 14 14 Wh...

Страница 31: ...f the product to the factory The sole responsibility of Koehler Instrument Company and the purchaser s exclusive remedy for any claim arising out of the purchase of any product is the repair or replac...

Страница 32: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 28 Notes...

Страница 33: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 29 Notes...

Страница 34: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 30 Notes...

Отзывы: