KV5000 Instrument & Software
Operation and Instruction Manual
KV5000-Manual
-
20-
Press the Viscometer Data Box (section 5.3, point
3) to access the Viscometer Setup Screen. See
Figure 4 in section 5.3 and Figure 6 below for an
additional illustration.
Figure 4. Control Panel_Viscometer Setup Screen
Refer to Section 5.3, points 5 thru 12 for
instructions on how to input and load viscometer
parameter data.
After all required parameters are entered and the
sample temperature has reached the test
temperature it is time to begin the test.
For Automatic Operation:
From the Main Screen, press the
Auto Mode
button.
Press the Start button.
. The viscometer
status field located at the upper right hand side of
the screen will change from
OFF
to
READY
.
Remove the stopper. When the sample passes the
first indication line of the viscometer tube the
instrument status will automatically change to
RUN
and the integrated timer will begin.
When the sample passes the second indication
line the time will automatically stop, the status will
display
DONE
, and a viscosity value will
automatically be displayed at the bottom of the
Test Data area (Figure 3, point 4).
Pressing and holding the Stop button
for 2
seconds or more will reset the time and viscosity
values and OFF will display as the status.
When using a viscometer with two (2) constants,
two (2) time values are used to calculate viscosity.
When the sample passes the first indicator line the
Time 1
will stop and
Time 2
will begin. The status
will automatically change from
RUN
to
RUN 2
.
Once the sample passes the second indicator line
on the viscometer tube the
TIME 2
will
automatically stop, the status will change from
RUN 2
to
DONE
and the viscosity value will
automatically be calculated and displayed.
For Manual Operation:
From the Main Screen, press the
Manual Mode
button.
Remove the stopper. When the sample passes the
first indication line of the viscometer tube, press
the Start
button
on the Control Panel. When
the viscometer is in standby mode the status will
display
OFF
at the upper right hand side of the
screen. Upon pressing the Start button
the
instrument status will change to
RUN
and the
integrated timer will begin.
When the sample passes the second indication
line, press the Stop button
. The time will
stop, the status will display
DONE
, and a viscosity
value will automatically be displayed at the bottom
of the Test Data area (Figure 3, point 4).
Pressing and holding the Stop button
for 2
seconds or more will reset the time and viscosity
values and OFF will display as the status.
When using a viscometer with two (2) constants,
two (2) time values are used to calculate viscosity.
This means that when running a viscosity test
using a two constant viscometer the Stop button
must be pressed twice.
When the sample passes the first indicator line the
Time 1
will stop and
Time 2
will begin. The status
will change from
RUN
to
RUN 2
.
Once the sample passes the second indicator line
on the viscometer tube, press the Stop button
a second time and the
TIME 2
will stop. The
status will change from
RUN 2
to
DONE
and the
viscosity value will automatically be calculated and
displayed.
Содержание KV5000
Страница 2: ......
Страница 4: ......
Страница 5: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 1...
Страница 6: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 2...
Страница 29: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 25 9 Wiring Diagrams 9 1 Unit Wiring...
Страница 32: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 28 Notes...
Страница 33: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 29 Notes...
Страница 34: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 30 Notes...