KV5000 Instrument & Software
Operation and Instruction Manual
KV5000-Manual
-
8-
DIN 51550: Determination of Kinematic
Viscosity and Dynamic Viscosity
5. Federal Test Method (FTM)
FTM Publication:
FTM 791-305: Kinematic Viscosity of
Petroleum Products
6. Association
Française
de
Normalisation
(AFNOR)
http://www.afnor.fr
AFNOR Publication:
NF T 60-100: Kinematic Viscosity of
Petroleum Products
1.3 Instrument Specifications
Models:
K23702-OS, K23708-OS,
K23792-OS, K23798-OS
Electrical
Requirements:
115V 60Hz
220V 50/60Hz
Temperature Range:
Ambient to 150°C
(302°F)
Optional Temperature
Range:
-20° to 150°C
(-4° to 302°F)
(Requires an optional
external chiller)
Temperature Control
Stability:
Exceeds ASTM
requirements
Viscometer Ports:
Seven (7) round 2”
(51mm) ports
Capacity:
Five (5) optical sensors
Bath Medium:
Water or suitable heat
transfer fluid
Bath Medium
Capacity:
22L (5.8 gal) or 34L (8.9
gal)
2 Safety Information and Warnings
Safety
Considerations.
The
use
of
this
equipment may involve
hazardous
materials and
operations. This manual does not purport to
address all of the safety problems associated with
the use of this equipment. It is the responsibility of
any user of this equipment to investigate,
research, and establish appropriate safety and
health practices and determine the applicability of
regulatory limitations prior to use.
Equipment Modifications and Replacement
Parts.
Any modification or alteration of this
equipment from that of factory specifications is not
recommended voids the manufacturer warranty,
product safety, performance specifications, and/or
certifications whether specified or implied, and
may result in personal injury and/or property loss.
Replacement parts must be O.E.M. exact
replacement equipment.
Unit Design.
This equipment is specifically
designed for use in accordance with the applicable
standard test methods listed in section 1.2 of this
manual. The use of this equipment in accordance
with any other test procedures, or for any other
purpose, is not recommended and may be
extremely hazardous.
Over Temperature Protection.
This unit is
equipped with Over Temperature Protection (OTP)
circuitry to prevent overheating. The unit will
automatically interrupt power whether equipment
malfunction
or
operator
error
causes
the
temperature to exceed either 20 °C above the set
point or the maximum recommended temperature
range. The power can only then be restored by
identifying and correcting the problem, allowing
the
unit
to
return
to
normal
operating
temperatures, and resetting the power to the unit.
Chemical Reagents Information.
Chemicals and
reagents used in performing the test may exhibit
potential hazards. Any user must be familiarized
with the possible dangers before use. We also
recommend consulting the Material Data and
Safety Sheet (MSDS) on each chemical reagent
for additional information. MSDS information can
be
easily
located
on
the
internet
at
http://siri.uvm.edu
or
http://www.sigma-
aldrich.com.
Содержание KV5000
Страница 2: ......
Страница 4: ......
Страница 5: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 1...
Страница 6: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 2...
Страница 29: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 25 9 Wiring Diagrams 9 1 Unit Wiring...
Страница 32: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 28 Notes...
Страница 33: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 29 Notes...
Страница 34: ...KV5000 Instrument Software Operation and Instruction Manual KV5000 Manual 30 Notes...