25
Removing the Drive Unit
1) Move probe into SERVICE position.
Only in this position can the drive unit be removed.
2) Switch off compressed air and deaerate device!
3) Make sure that no process fluid is leaking from the outlet.
4) Turn coupling nut counterclockwise using the ZU 0648 mounting
wrench (see “Accessories”, page 35). This pulls the drive unit
out of the process unit.
You can slightly lift up the drive unit while turning the nut to
support its movement. Do not cant the unit and do not exert
force. Do not loosen the coupling nut completely.
5) Make sure that no process fluid is leaking from the outlet.
6) Completely loosen the coupling nut and pull off the drive unit
upwards.
ZU 0648
NOTICE
Be sure to follow these instructions in the correct sequence!
Take appropriate safety precautions against escaping process
fluids.