13
Information on Installation
For operation of the Ceramat, you must connect control air, rinsing or calibration media, and the
electrical check-back signal for indicating the probe position.
There are two basic options for this.
When the Ceramat is operated with a Unical or Uniclean electro-pneumatic controller and the Protos measur-
ing system, the cables and tubings for air pressure, rinsing/calibration media and check-back are combined in
a single hose with just one plug connection (multiplug). This hose is referred to as media connection.
This media connection is installed on the Ceramat together with the outlet hose.
When you do not use a probe controller (Unical or Uniclean and Protos measuring system), you can connect
the supply lines for control air, rinsing/calibration media and electrical check-back to the Ceramat with a free
hose connection via the ZU 0631 standard-media interface.
Connection for operation with
Unical/Uniclean und Protos
Connection for operation with
ZU 0631 standard media interface
Содержание Ceramat WA 153
Страница 1: ...User Manual English Latest Product Information www knick de Ceramat WA 153 Sensor Lock Gate ...
Страница 27: ...27 Installation Dimensions Approx 600 mm clearance for sensor installation Solid Electrolyte ...
Страница 28: ...28 Installation Dimensions Approx 800 mm clearance for sensor installation Liquid Electrolyte ...
Страница 42: ......
Страница 43: ......