14
Information on Installation
15°
15°
Installing the outlet hose
Do not install the outlet hose more than 1 meter below calibration chamber level (see figure 2).
Doing so raises the danger of air being sucked out of the calibration chamber due to the drop in pressure.
Leakage of the calibration chamber due to gravity alone is prevented by the loop in the outlet hose when the
Ceramat is installed at an angle of up to 15° above the horizontal (see figure 1).
With horizontal to upside-down installations, you must install an appropriate hose bend above the level of
the calibration chamber to prevent leakage from the calibration chamber (see figure 2).
max.
1 meter
Calibration chamber
level
Above level of calibration chamber
Calibration chamber
Outlet hose
Hose bend
Inlet
Sensor
Fig. 1
Fig. 2