The Ceramat pneumatic sensor lock-gate with ceramic sealing is used for installing a sensor for measuring
process variables in liquid media. The sensor can be cleaned, calibrated, or replaced under process conditions
(pressure and temperature).
Since the pneumatic drive unit and the process unit with the ceramic sensor lock-gate are separate modules,
the drive unit can be removed and replaced under process conditions.
The sensor lock-gate is suitable for sensors with an outer diameter of 12 mm, for sensors:
• with solid electrolyte, length 225 mm, sensor head with PG 13.5
• with liquid electrolyte, length 250 mm
• for optical sensors with a body diameter of 12 mm or 12.7 mm (1/2“)
(For installing an optical sensor, you require a special adapter which is listed in the Product Coding under
“Special version”. Information on design and use of the suitable adapter is provided in a supplementary
datasheet.)
The Ceramat sensor lock-gate allows:
• inserting and retracting the sensor under process pressure (sensor lock-gate)
• calibrating or adjusting the measuring system and cleaning the sensor in the running process
(different options available)
• removing or replacing the sensor after separation from the process (in SERVICE position)
The Ceramat sensor lock-gate has been developed and manufactured in compliance with the applicable
European directives and standards.
Compliance with the European directives and standards is confirmed by the EC Declaration of Conformity.
Compliance with the European Harmonized Standards for use in hazardous locations is confirmed by the
EC-Type-Examination Certificate.
To ensure smooth performance of the measuring points with the Ceramat, a number of important operating
and maintenance parameters must be complied with (see Ceramat Maintenance Instructions).
There is no particular direct hazard caused by the operation of the device in the specified environment.
Intended Use
!
CAUTION
Safe use
• If you are not sure whether the sensor lock-gate can be safely used for your intended application, please
contact the manufacturer.
• Take account of the influences of humidity, ambient temperature, chemicals and corrosion.
• To ensure safe use of the equipment, you must follow the instructions given in this manual and observe
the specified temperature and pressure ranges.
!
WARNING
Application in hazardous locations
For hazardous-area applications, the sensors used must ensure proper separation of the ATEX zones. When
the sensor lock-gate is in SERVICE position, the sensor may be replaced within a Zone 1 hazardous location.
Содержание Ceramat WA 153
Страница 1: ...User Manual English Latest Product Information www knick de Ceramat WA 153 Sensor Lock Gate ...
Страница 27: ...27 Installation Dimensions Approx 600 mm clearance for sensor installation Solid Electrolyte ...
Страница 28: ...28 Installation Dimensions Approx 800 mm clearance for sensor installation Liquid Electrolyte ...
Страница 42: ......
Страница 43: ......