30
Connecting the Smartline Pump 1000 to other Instruments
Fig. 23
Mounted SmartMix at the pumps front pannel
Depending on the use in a HPG or LPG system opposit flow directions
are necessary. For this purpose the mixer shall be turned by 180°, see
Fig. 24.
Injector
HPG LPG
Pump 1
Pump 2
Pump
Fig. 24
System depending orientation of the SmartMix mixer
SOP 22
Connecting the SmartMix mixer in a LPG system
1. Mount the SmartMix mixer as shown in Fig. 23.
2. Connect the outlet of the pressure transducer with the single inlet
at the bottom of the mixer (Fig. 24).
3. Connect one of the mixers outlets with the inlet of the injection
valve.
4. The second outlet is to be left closed with a blind fitting.
SOP 23
Connecting the SmartMix mixer in a HPG system
1. Mount the SmartMix mixer turned by 180° compared to Fig. 23.
2. Connect the outlets of the pressure transducers of both pumps to
the inlets of the mixer cover, now placed at the bottom (Fig. 24).
3. Connect the mixers outlet with the inlet of the injection valve.