Control of the Smartline Pump 1000
11
Links:
21
___
___
___
___
(21 . . 29)
___
___
___
___
Load
Edit Prog
Flow
Pressure
Events
Prog
Clear 0
[ml/min]
[0.1 Mpa]
off 0 on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
rhombus
Fig. 8
Example of a multi-page menu: VIEW menu
Menus without a rhombus
in the lower left corner of the display are
single page menus.
SOP 2
Entering Data
This SOP applies to the Smartline Pump 1000 Firmware Version 1.0X or
higher. Use this SOP for entering parameters and program steps in your
Smartline Pump 1000.
1. Place the cursor at the desired position of the main screen.
2. Press a numeric key to start data input. The appearance of the
cursor changes to a underscore character (see Fig. 6 on page 10).
3. Press the numeric keys to enter the desired numeric values.
Incorrect entries can be simply overwritten by continuing the entry.
4. Confirm the value by pressing any of the arrow keys. This finishes
data entry mode and moves the cursor to the next field according
to the arrow key direction.
5. You can modify and delete data according to SOP 15 “Modifying
and Deleting Programs” on page 24.
Using the Arrow keys in the SETUP Menu
Fig. 9 “Structure of the SETUP Menu” on page 12 gives a complete
overview of the Setup Menu.
You can use the
¾
arrow key for accessing menus.
You can use the
¿
and
À
arrow keys for scrolling through the list
of parameters of any individual menu item.