Programming the Smartline Pump 1000
23
01
07
001.0
|200| 0 0 0 0 0 0 0 0
OFF
000.00
100
0
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
A
01
07
001.0
|200|
0 0 0 0 0 0 0 0
OFF
000.00
90
10
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
B
01
07
- - -.-
|200| -- -- -- -- -- -- -- --
OFF
***.*
- -
- -
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
C
01
07
_ _ _._
|200|
_ _ _ _ _ _ _ _
OFF
020.0
10
90
Load
Edit Prog
Flow
Pressure Events
Prog
Clear 0
[ml/min]
[0.1 Mpa] off 0
on 1
2
Hold 0
Time
%A
%B
%C
%D
Run 1
[min]
Setup
GLP
Link
View
D
Fig. 17
Display of single steps, when creating a program
(see also “Example for creating a program” on page 25)
SOP 14
Execution of Programs
This SOP applies to the Smartline Pump 1000 Firmware Version 1.0X or
higher.
1. Load the program you want to execute by entering the program
number in the LOAD PROG field. Confirm by pressing any arrow key.
2. Press START key to start the pump. The display changes to
ON
and
the pump begins continuous operation with the parameters specified
in the first program line.
3. Position the cursor onto the
ON
field.
4. Start the program by pressing the numeric key 1. The display will
show that the program is running by indicating
RUN
on the lower left
side of the display.
5. While a sequence of programs is being executed, it is possible to
freeze the current status of the pump by pressing the 0 key. The
display will indicate
HOLD
in the lower left corner and the TIME field
will blink.
6. By pressing again on the 1 key the active program sequence will
continue running again. You can also press
STOP
to abort execution.