Unpacking and setup
16
AZURA® Valve Unifier VU 4.1 Instructions V6855
Binary control
If the valve drive was set to binary operation (see “Input” on p. 26), then the
connections BIN 0 - BIN 3 are available as inputs.
Binary code
A binary code is entered during binary control so that the valve can be set
externally in the correct position (nominal position).
HOME
TTL input
Low active
Secure switching threshold min. 10 mA
After receiving a signal (short-circuit to GND) from an exter-
nal device:
Valve drive is set to position 1.
BACKWARD/
INJECT
TTL input
Low active
Secure switching threshold min. 10 mA
After receiving a signal (short-circuit to GND) from an exter-
nal device:
INJECT (position 2 for 2 position valves)
Move to the next lower port of the valve, e.g. from posi-
tion 6 to position 5
FORWARD/
LOAD
TTL input
Low active
Secure switching threshold min. 10 mA
After receiving a signal (short-circuit to GND) from an exter-
nal device:
LOAD (position 1 for 2 position valves)
Move to the next higher port of the valve, e.g. from posi-
tion 2 to position 3
START IN
TTL input
Low active
Secure switching threshold min. 10 mA
After receiving a signal (short-circuit to GND) from an exter-
nal device, the device starts.
Signal
Explanation