Unpacking and setup
12
AZURA® Valve Unifier VU 4.1 Instructions V6855
1
and a computer
5
. To avoid interference, we recommend operating the
HPLC system separately from the company network.
You will find patch cables for each device and the router in the accessories kit.
To connect the router to a WAN, an additional patch cable is required, which
is not supplied within the scope of delivery.
Prerequisite
The computer has been switched off.
There is a patch cable for each device and the computer.
Process
1. Use the patch cable to connect the router and the computer. Repeat this
step to connect all devices.
2. Use the power supply to connect the router to the mains power system.
Configuring the router
The router is preset at the factory. The login information is mentioned on the
router case (IP address, user name, and password), which is needed for
router configuration.
Process
1. To open the router configuration, start your Internet browser and enter the
IP address (not for all routers).
2. Enter user name and password.
3. Configure the router as DHCP server.
4. In the router configuration, check the IP address range and make changes
if necessary.
Note: If the IP address range has been changed, it is necessary to note it
down.
1
2
3
4 5