9.4
Cleaning Process
Due to the better results regarding efficacy and repeatability, machine cleaning should be
preferred to manual cleaning.
Description
REF
Tip
Tube
Focusing handpiece 50 mm
76-100-06-04
76-100-07-04
Focusing handpiece 127 mm
76-100-11-04
76-100-12-04
Focusing handpiece 200 mm
76-100-16-04
76-100-17-04
Scanner handpiece 127 mm
76-500-51-04
76-500-52-04
Scanner handpiece 200 mm
76-500-61-04
76-500-62-04
Backstop 127 mm
76-100-20-04
Backstop 200 mm
76-100-25-04
Adapter
76-200-15-04
Angle 120°
76-200-10-04
Angle 90°
76-200-05-04
ENT handpiece tube
76-600-11-04
ENT attachment 0°
76-600-02-04
ENT attachment 90°
76-600-03-04
ENT attachment 120°
76-600-04-04
Manual precleaning
Immerse into cold water for 5 min
Automatic cleaning
Cleaning and disinfection machine
acc. to EN ISO 15883-1
e.g. RDA G7735 CD/Miele
Cleaning program
Vario TD
Positioning
MIC tray, preferably in own sieve tray
Pre-washing
2 min in cold water
Type of cleaning agent
alkaline
Possible cleaning agents
Neodisher
®
MediClean forte 0.5%
Temperature
+55°C (+131°F)
Cleaning time
5 min
Neutralizing
3 min in cold water
Rinsing
2 min in cold water