![KLS Martin MCO 25 plus Скачать руководство пользователя страница 66](http://html1.mh-extra.com/html/kls-martin/mco-25-plus/mco-25-plus_instructions-for-use-manual_1979006066.webp)
66
Revision 5.5
The order in which the laser dots are generated is irregular, which minimizes the thermal
strain on the tissue.
Fig. 6-9:
Generation of individual dots in irregular order (example: square)
6.7.4
POWER FRAC
For a dotted scanning effect with high penetration depth, connect the scanner handpiece
POWER FRAC
to the scanner and select the setting
POWER FRAC
in the menu
HAND PIECE
. When using the scanner, the laser generates a surface-filling number of spots
with a diameter of 0.2 mm.
Setting POWER FRAC, example scan shape square
Handpiece selection
POWER FRAC
Setting ranges
POWER
10–20 W
FIGURE SIZE
1–5 mm
SCAN SHAPES
Triangle, rectangle, polygon
FILL FACTOR
LIGHT
20% of area filled
MEDIUM
50% of area filled
STRONG
80% of area filled
Energy density
20–39 J/cm²
Energy density
The fill factor determines the number of fractionated laser spots on the selected scan shape.