![KLS Martin MCO 25 plus Скачать руководство пользователя страница 116](http://html1.mh-extra.com/html/kls-martin/mco-25-plus/mco-25-plus_instructions-for-use-manual_1979006116.webp)
116
Revision 5.5
Recommended separation distances between portable and mobile RF telecommunications
equipment and the Gebrüder Martin MCO 25 plus according to IEC 60601-1-2
clause 6.8.3.201 b), Table 206
The device MCO 25 plus is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF telecommunications equipment (transmitters) and
the device as recommended below, depending on the maximum output power of the communications
equipment.
Separation distances according to frequency of transmitter (in m)
Rated maximum
output
P
of
transmitter
in watts
150 kHz to 80 MHz
P
d
17
.
1
=
80 MHz to 800 MHz
P
d
17
.
1
=
800 MHz to 2.5 GHz
P
d
33
.
2
=
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.7
3.7
7.4
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where
P
is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
NOTICE
Portable and mobile RF communication equipment can interfere with the performance of
the CO
2
Laser System MCO 25 plus.
•
Observe “
”, see table!