Revision 5.5
•
Press the key
LASER READY
(S3) to set the laser to readiness.
The MCO 25 plus laser now internally checks the selected laser power.
During this check, the lamps of the keys
STANDBY
(S1) and
LASER READY
(S3) flash in alternating fashion.
•
After the check, the key
LASER READY
(S3) is permanently lit yellow.
CAUTION
Danger of physical injury due to wrong parameters!
Before activating the foot switch on the control panel, check the set parameters once more.
•
Switch on the laser light by pressing the foot switch.
—
The laser emission display
L1
is lit.
—
At the same time, a high warning sound can be heard for the duration of the laser
treatment.
If the foot switch is released before the set pulse time has elapsed, the laser emission stops
immediately. The originally set pulse time will be used again if the foot switch is pressed again.