Revision 5.5
3.9
Laser Safety Officer
According to the German Accident Prevention Directive BGV B2 (Laser Radiation), the operator
must appoint a laser safety officer in writing. This person is also responsible for the following:
•
Implementation of safety measures
•
Instruction of all persons working with the device in the safety measures
•
Instruction of all persons using the device in the operation of the device
•
Marking of the laser area with signs
•
Inspection of the warning signals
•
Safe storage of the keys
•
Correct connection of the device in case of location changes
•
Keeping of the medical product log
3.10
User
The user is responsible for safe and proper use of the medical laser system. He or she must
ensure that all safety precautions are complied with, and be familiar with the laser-specific
surgical technique.
3.11
Electrical Safety
The device conforms to protection class 1 (IEC 60601), which means that it must be connected
to a properly earthed mains according to its technical data, see section 13 “Technical Data”,
page 109.
•
Use only the mains cable included in the delivery.
•
Mains cable and plugs must not show any signs of damage.
•
The laser system must never be activated if any part of the casing is not attached
correctly. If the casing is not properly closed, laser radiation may escape uncontrollably,
and dangerous electrical voltages or currents may be exposed.
3.12
Grounding the Unit
The unit is grounded via the ground conductor integrated in the mains cable. This is an
important prerequisite for the safe operation of the unit. If the mains cable is correctly
connected in accordance with relevant electrotechnical regulations, proper grounding is
automatically ensured.
The yellow-and-green equipotential bonding cable is intended for use in cases where potential
equalization is required for laser operation (e.g. when using the Limax
®
120 for intracardiac
interventions).