Revision 5.5
5.7.1
Mounting and Securing the Mirror Joint Arm
NOTICE
Remove latch (4, Fig. 5-2) for removing or mounting the mirror joint arm!
•
Remove the cap (7) from the mirror joint arm.
•
Screw mirror joint arm into the connection thread (1, Fig. 5-2) on the laser surgery device.
Fig. 5-4:
Overview of mounting process
1 Lower part of the arm (to be plugged
into the laser surgery device)
2 Arm latch (
locked/unlocked
)
Working position
locked
: Arm remains in
upright position
3 Setting for spring tension:
(
handpiece
/
scanner
)
normal spring force =
handpiece
strong spring force =
scanner
4 Setting screw for rotary fixation of the
mirror joint arm
5 Elastic arm holder
6 Cover ring
7 Protective cap on mounting thread
8 Laser surgery device retainer
9 Air outlet for flushing air hose
•
Loosen setting screw (4).
•
Insert arm with thread in vertical position into the designated retainer (8) on the laser
surgery device.