Status-Code
Error
description
First step done
by user
Second step
done by user
Service
operation done
by technician
9007
Critical error:
System not
calibrated while
calibrating chart
is accessed
Power off and on
Call service
Run new
software and
calibrate
9012
Critical error:
Internal
operating hour
software error
Power off and on
Call service
Run new
software and
calibrate
9013
Critical error:
Internal laser
state software
error
Power off and on
Call service
Run new
software and
calibrate
9014
Critical error:
Internal timer
software error
Power off and on
Call service
Run new
software and
calibrate
9015
Critical error:
Controller
system self
deactivated
Power off and on
Call service
Run new
software and
calibrate
9016
Critical error:
Internal
calibrating
software error
Power off and on
Call service
Run new
software and
calibrate
9101
Critical error:
Atmel bootloader
error (protocol
error)
Power off and on
call service
Run new
software and
calibrate
9102
Critical error:
Atmel bootloader
error (memory
or transmission
error)
Power off and on
call service
Run new
software and
calibrate
9103
Critical error:
Atmel bootloader
error (error while
ending
transmission)
Power off and on
call service
Run new
software and
calibrate