![KLS Martin LIMAX 120 Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/kls-martin/limax-120/limax-120_service-manual_1979001077.webp)
Service manual LIMAX
120
Service manual version 1.00
page 74
Status-Code
Error
description
First step done
by user
Second step
done by user
Service
operation done
by technician
2007
Critical error:
5 V supply
voltage error
Power off and on
Call service
Replace
controller PCB or
power supply
unit
2008
Critical error:
Power supply
unit general
error
Power off and on
Call service
Replace power
supply unit
2009
Critical error:
Controller Self-
test
communication
error
Power off and on
Call service
Replace
controller PCB
2010
Critical error:
safety system
internal error
Power off and on
Call service
Replace PCB
2011
Critical error:
-12V supply
voltage failure
Power off and on
Call service
Replace power
supply unit or
controller board
2013
Grave error:
controller system
error
Press down
STANDBY
Call service
Replace
controller PCB
2014
Grave error:
controller system
communication
error
Press down
STANDBY
Call service
Replace
controller PCB
2015
Grave error:
controller system
communication
error
Press down
STANDBY
Call service
Replace
controller PCB
2016
Grave error:
Safety shutter
not closed
Press down
STANDBY
Call service
Replace
controller PCB or
power supply
unit