Service manual LIMAX
120
Service manual version 1.00
page 59
Status codes VAC
Status-
Code
Error description
Since version
Internal errors or logic errors
1
Critical error: undefined error
Measure(s): contact support
v0.87
2
Warning: a coding error occurs
Measure(s): no error indication is effected on device;
service/MCB master can read out the error memory and inform
the support
v0.87
3
Critical error: maximum number of error definitions is
exceeded(coding error exception)
Measure(s): contact support
v0.87
4
Critical error: internal error list is inconsistent and corrupt
(coding error exception)
Measure(s): contact support
v0.87
5
Warning: nullpointer occurred
Measure(s): no error indication is effected on device;
service/MCB master can read out the error memory and inform
the support
v0.87
6
Severe error: a flow parameter is corrupt (hardware or
software error)
Measure(s): firmware update if the error is known (and if it
occurred at other devices under the same conditions), otherwise
change the motherboard
v0.441
Status-
Code
Error description
Since version
Watchdog
50
Critical error: watchdog-reset was operated based on
abnormal program situations
Measure(s): contact support
v0.138
51
Critical error: watchdog-reset wasn’t operated while
watchdog-test
Measure(s): contact support
v0.138
52
Critical error: frequency of the main control is incorrect
Measure(s): check FUSE-Bits CKSEL (ext. Crystal Osc. 8 MHz)
or rather check dimensioning of the external oscillator
(16MHz), contact support
v0.138