6
Filling the water cooling unit
6.1
First water fill in
Remove the front cover (chapter 5.1) and turn on the laser. Open the membrane element at
the end of the air-vent tube by turning it counter clockwise by approximately 360°.
Fig. 6.1 Air-vent tube sealed (left picture) and slightly opened (right picture)
Thus the air can quickly get out during the filling procedure. Now fill in the first litre of de-
ionized water into the cooling circuit. Therefore, remove the stainless steel set into the filling
tube. Then press down the clamp collar to get out the plug. Use the plastic wrench from the
service tool kit, to push down the clamping mechanism. It can also be done by hand.
The filling tube is seen on the right side of the opened laser front.
Fig 6.2 Front of LIMAX
®
120 with filling tube, removing of the plug with wrench.