Service manual LIMAX
120
Service manual version 1.00
page 33
5.4
Removal of the gas flow controller
Shut down the LIMAX
®
120 laser unit and remove the backside cover (chapter 5.2). Remove
the air inlet and outlet connection tubes (2 and 3) from the gas flow controller and the valve
connector cables (4) from the PCB stack. Remove three M4x25 socket head cap screws (1) the
distance sleeves (5, fig. 5.6a) using a 3 mm hex screw driver.
Previous version
current version
Fig. 5.6 Removing the gas flow controller
1
M4x20 socket head cap screw
3 air outlet tube
2
air inlet tube
4 valve connector cables
Fig. 5.6a distance sleeves
5
5
5