Service manual LIMAX
120
Service manual version 1.00
page 44
6.3
Second step to fill in the de-ionized water
Start the LIMAX
®
120 laser unit. The water pump is running. Go into the service menu by
pressing the rotary switch and select the menu item“> Setup” and “> Service”.
Type in the service pin “2-0-2-0” and select “> “Cooling”.
Check the parameter: “Filling level” can have 3 values:
Filling level: LOW not enough water – water pump is switched off for safety
Filling level: WARN enough water that the pump can run – refill water
Filling level: OK
well filled water circuit
Following chapter 6.1 the filling level should indicate WARN and the water pump should be
working.
Now fill in additional water (approximately 2 litres) until the filling level shows: OK
Fill additional water until the water level is in the middle of the filling tube. Seal the filling tube
with the plug.
Fig. 6.4