![KLS Martin LIMAX 120 Скачать руководство пользователя страница 70](http://html1.mh-extra.com/html/kls-martin/limax-120/limax-120_service-manual_1979001070.webp)
Service manual LIMAX
120
Service manual version 1.00
page 67
11
LIMAX
®
120 status codes
The different status codes are classified in 5 groups:
Severe error: The status code is displayed, the laser as well as the additional components is
shut down and an entry is made into the service-log. Even after rebooting the LIMAX the laser
can not be reactivated as long as a service technician did not log into the SERVICE menu item.
Critical error: The status code is displayed, the laser as well as the additional components is
shut down and an entry is made into the service-log. The LIMAX has to be rebooted to activate
the laser or the additional components.
Grave error: The status code is displayed, the laser as well as the additional components is
shut down and an entry is made into the service-log. The laser can be reactivated as soon as
the STANDBY key is pressed down.
Minor error: The status code is displayed, the laser as well as the additional components is
shut down and an entry is made into the service-log. After a short period the status code
disappears and the laser can be reactivated.
Warning: An entry is made into the service log. The laser operation is not interrupted by the
warning.
Status-Code
Error
description
First step done
by user
Second step
done by user
Service
operation done
by technician
1001
Critical error:
Memory error
Power off and on Call service
Replace
controller PCB
1002
Critical error:
Footswitch
operated during
start up
Power off and on
Don’t operate
the footswitch
while powering
on
Check if
footswitch
jammed. Test/
replace
footswitch
1004
Critical error:
Self-test: safety
relais failed to
open
Power off and on
call service
Test/ replace
power supply
unit or controller
PCB
1005
Critical error:
Self-test: safety
relais failed to
close
Power off and on
call service
Test/ replace
power supply
unit or controller
PCB
1008
Minor error:
Overcurrent on
diode module
Wait 30 s before
reactivating
laser
call service
Test/ replace
power supply
unit