![KLS Martin LIMAX 120 Скачать руководство пользователя страница 74](http://html1.mh-extra.com/html/kls-martin/limax-120/limax-120_service-manual_1979001074.webp)
Service manual LIMAX
120
Service manual version 1.00
page 71
Status-Code
Error
description
First step done
by user
Second step
done by user
Service
operation done
by technician
1048
Critical error:
Atmel selftest
error
Power off and on
Call service
Replace
controller PCB
1049
Warning: high
pressure in
cooling unit
Call service
-
Test/Replace
cooling unit
1050
Critical error:
UH4 auxiliary
voltage failure
Power off and on
Call service
Replace power
supply unit or
controller PCB
1051
Critical error:
Footswitch
contactors 1st.
switching stage
error
Power off and on
Call service
Replace
Footswitch or
power supply
unit
1052
Critical error:
Footswitch
contactors 2nd.
switching stage
error
Power off and on
Call service
Replace
Footswitch or
power supply
unit
1053
Critical error:
Footswitch SWAP
contactors error
Power off and on
Call service
Replace
Footswitch or
power supply
unit
1054
Critical error:
DI-water Temp.
low
Let LIMAX warm
up then Power
off and on
Call service
Replace cooling
unit
1055
Critical error:
DI-water Temp.
high
Let LIMAX cool
down then Power
off and on
Call service
Replace cooling
unit
1056
Severe error:
Laser fiber
Temp. high
Call service
-
Test Laser/
Replace
Hardware
1057
Severe error:
Laser disabled
after code 1056
Call service
-
Reset Error with
service PIN