5.9
Removal/ Replace of the particle filter
Always replace the water in the cooling system when replacing the particle filter.
Shut down the LIMAX
®
120 laser unit. Remove the front cover (chapter 5.1), the back cover
(chapter 5.2) and drain the cooling water (chapter 7). In addition open the
sealing plug at the top side of the laser resonator (fig 6.4), remove the particle
filter by unscrewing the cap (2,
Fig. 5.17) and then remove the filter (1). Put in the new particle filter, reconnect the
connectors and fill the water cooling unit (chapter 6).
Fig. 5.17 Removal of the particle filter
1
Particle filter
2
Filter cap