Service manual LIMAX
120
Service manual version 1.00
page 58
9.5
VAC
Enter the menu item VAC by selecting the menu item and push down the rotary switch.
Fig. 9.5 Enter the VAC menu item
In the VAC menu all important information and settings are listed. The versions give
information about the state of hardware and software (1, fig. 9.5a). Settings of work and basic
suction, which can be set in the ordinary VAC menu, are displayed as well as their bounds (2
and 3). Furthermore information is given about the VAC status, possible error codes (table 1)
and the temperature of the blower (4,5 and 6). The filter of the VAC should be replaced after
40 operating hours. To recognize when the filter has to be changed there is a percentage value
that indicates the remaining capacity (7). The adhere detect is set on 0% from the factory (8).
That means that sensitivity is completely disabled. If the suction device touches the tissue, the
VAC neither interrupts nor stops. The opposite would be the case if the adhere detect is set on
a high value. Depending on sensitivity the VAC will stop activation immediately if the suction
device contacts tissue.
Fig. 9.5a VAC menu