7.2
Draining for temperature > 0 °C, frost free
Shut down the LIMAX
®
120 laser unit. Remove the front cover (chapter 5.1).
Fig 7.1
1 and 2 draining tube
3 particle filter
Draining tube 1 is for emptying the heat exchanger and tube 2 is for emptying the tank and
pump.
Open the sealing plugs of draining tubes 1 and 2, bend the connectors down, and collect the
water in a flask.
The leaking water is producing a negative pressure, so the connected tubes are drained too. It
could be helpful to open the sealing plug at the filling tube. Open the particle filter to let out
the water inside.