PUSH ON OPTICAL CONNECTOR
KI23400/KI27400 UM-2
-
21 -
If this is not enough, then clean the interface by rubbing a lint-free lens
cloth over the surface using small circular movements.
Connector Wear: Fiber Optic Connectors usually have a design life of
around 1,000 insertions before they are worn out.
So, if your instrument has had considerable use, the removable
connector adaptor may be worn out, and need replacing.
If your source has become very unstable, the through connector ceramic
sleeve may be cracked, and the push-on through connector needs
replacing.
It is good practice to use test leads that can be easily replaced as
needed.
When performing precision measurements, it is usually helpful to have
an APC connector plugged into a power meter, to reduce line reflections.
CAUTION!
Disable the source when cleaning optical interfaces.
Cleaning materials and chemicals can react with the intense emitted
light and, damage the connector.
When To Clean The Optical Connectors
Every time they are mated. You will need a ready supply of suitable
connector cleaning and inspection equipment. This will not only help get
accurate readings but is also essential to protect your test equipment.