* Patient Warming Systems
M000 Control Unit Operator’s Manual: U .S .
Alarm/
Alert
Message
Displayed
Alarm/Alert
Description
System Status
Action
A/D Acquisition
Fault.
Alarm
A/D converter
fault.
Contact Kimberly-Clark Service.
•
Interprocessor
communication
failure (Monitor)
Alarm
The Monitor
Processor does
not communicate.
Contact Kimberly-Clark Service.
•
Interprocessor
communication
failure (Control)
Alarm
The Control
Processor does
not communicate.
Contact Kimberly-Clark Service.
•
Machine Type
Fault.
Alarm
Unit has very little
or no cleaning
solution in the
reservoir.
Turn unit off, drain unit, fill unit using cleaning
kit (part #00838-01). See cleaning kit instructions
for details.
Contact Kimberly-Clark Service.
•
•
Power Supply
Fault.
Alarm
Power supply volt-
age out of range.
Contact Kimberly-Clark Service.
•
Processor
Fault.
Alarm
Processor Fault.
Contact Kimberly-Clark Service.
•
Unexpected
interrupt.
Alarm
Processor fault.
Contact Kimberly-Clark Service.
•
Processor
Fault.
Alarm
Processor Fault.
Contact Kimberly-Clark Service.
•
82
83
84
Appendix C — Alarm/Alert Table