Alarm/
Alert
Message
Displayed
Alarm/Alert
Description
System Status
Action
Water
temperature
high. Allow
system to
cool.
Alarm
Water temperature
is above 42.5°C
(108.5ºF).
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
This alarm may be caused by transient conditions
such as a momentary occlusion, air bolus, hot
restart, or no-load condition.
Check flow rate. Low flow rate or fluctuating flow
rates may cause overheating of the water.
Wait until water cools. Restart Automatic Mode
or Manual Mode.
If problem persists, contact Kimberly-Clark
Service.
•
•
•
•
Water
temperature
low. Allow
system to
warm.
Alarm
Water temperature
is below 3.5°C
(38.3ºF).
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
This alarm may be caused by transient conditions
such as a momentary occlusion, air bolus, hot
restart, or no-load condition.
Wait until water warms. Restart Automatic Mode
or Manual Mode.
If problem persists, contact Kimberly-Clark
Service.
•
•
•
Water
temperature
above safety
level. Allow
system to
cool.
Alarm
Water temperature
is below 44°C
(111.2ºF).
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
This alarm may be caused by transient conditions
such as a momentary occlusion, air bolus, hot
restart, or no-load condition.
Check flow rate. Low flow rate or fluctuating flow
rates may cause overheating of the water.
Wait until water cools. Restart Automatic Mode
or Manual Mode.
If problem persists, contact Kimberly-Clark
Service.
•
•
•
•
Water
temperature
below safety
level. Allow
system to
warm.
Alarm
Water temperature
is below 3.5°C
(38.3ºF).
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
This alarm may be caused by transient conditions
such as a momentary occlusion, air bolus, hot
restart, or no-load condition.
Wait until water warms. Restart Automatic Mode
or Manual Mode.
System can be run in Manual Mode, but Water
Target Temperature should be increased.
If problem persists, contact Kimberly-Clark
Service.
•
•
•
•
Water
temperature
below safety
level. Allow
system to
warm.
Alarm
Water temperature
is below 3°C
(37.4ºF).
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
This alarm may be caused by transient conditions
such as a momentary occlusion, air bolus, hot
restart, or no-load condition.
Wait until water warms. Restart Automatic Mode
or Manual Mode.
System can be run in Manual Mode, but Water
Target Temperature should be increased.
If problem persists, contact Kimberly-Clark
Service.
•
•
•
•
System
unable to
maintain
stable water
temperature.
Alarm
Alarm In Manual
Mode, system is
unable to
control water
within 1.0°C/F of
set point after 25
minutes in current
mode or at the
time of last change
in Water Target
Temperature.
Changes to Stop
Mode. Audio alarm
half-second tone
every 10 seconds.
Alarm/Alert
indicator flashing.
Contact Kimberly-Clark Service.
•