7
K
imberly
-c
larK
* Patient Warming Systems
M000 Control Unit Operator’s Manual: U .S .
The Patient Warming System M000 Control Unit is shipped with default settings
(see Tables 4-6 below), which may be modified as needed . Settings can be changed
for individual patients or procedures, or permanently changed to create new
default settings .
Prior to using the Patient Warming System on a patient, it is important to verify that
all parameters are appropriately set for the particular patient or procedure, including
maximum water temperature in Automatic Mode .
Table 4 indicates the parameters, the default settings, the available ranges of settings,
and incremental changes . When pressing the Down arrow, the parameters that are
displayed are shown in order of their appearance .
3 .5 Setting Custom
Parameters
Patient Target XX°C
Automatic Mode/
(Default-37°C) (98.6ºF)
Water Target XX°C
Manual Mode/
(Default-37°C) (98.6ºF)
Water Level
Maximum Water XX°C
Automatic Mode
(Default-42°C) (107.6ºF)
Minimum Water XX°C
Automatic Mode
(Default-4°C) (39.3ºF)
Set desired target
patient temperature in
Automatic Mode.
Set desired water
temperature in Manual
Mode.
Shows the level of
the water in the
reservoir at any given
time. This reading is
updated during purge,
fill cycle & power on.
Set the maximum
water temperature
the system will use
to control patient
temperature in
Automatic Mode.
Set the minimum
water temperature
the system will use
to control patient
temperature in
Automatic Mode.
33°C-37°C
(91.4ºF to 98.6ºF)
20°C-42°C
(68ºF to 105.8ºF)
Empty to Full
32°C-42°C
(89.6ºF to 107.6ºF)
20°C-32°C
(68ºF to 89.6ºF)
0.1°C (0.1ºF)
increments
1.0°C (1.0ºF)
increments
Empty, Low, Full
1.0°C (1.0ºF)
increments
1.0°C (1.0ºF)
increments
Parameter–Default
Explanation
Range
Resolution
Table 4
Setup provides a second tier of user-defined options that pertain to equipment
settings that may require less frequent changes . (See Tables 5 and 6)
Patient High Alert XX°C
Automatic Mode
(Default 42°C) (107.6ºF)
Patient Low Alert XXºC
Automatic Mode
(Default 10°C) (50ºF)
Defines the tempera-
ture limit at which the
system should issue
an alert if the patient
temperature exceeds
that limit.
Defines the tempera-
ture limit at which the
system should issue
an alert if the patient
temperature falls
below that limit.
10.1°C- 42.0°C
(50.1ºF to 107.6ºF)
10.0°C- 41.9°C
(50ºF to 107.5ºF)
0.1°C (0.1ºF)
0.1°C (0.1ºF)
Incremental
Parameter–Default
Explanation
Range
Changes
Table 5