M000 Control Unit Operator’s Manual: U .S .
The Patient Warming System M000 Control Unit can be operated from remote loca-
tions by attaching the User Interface (Figure 3) to an IV pole or other suitable holder .
The User Interface is connected to the Patient Warming System M000 by a .
metre (42 inch) cable . A 4 .6 metre (5 foot) cable is also available .
Figure 3
Kimberly
Clark
*
There are four principle modes of operation . These can be accessed by
pressing any of the four keys on the lower right side of the display: Automatic,
Manual, Purge, and Stop . Each key represents a unique function of the
KIMBERLY-CLARK* Patient Warming System M000 Control Unit .
a
utOmatic
m
Ode
:
The Automatic Mode monitors and controls the patient’s temperature to a set target
temperature . Water will be cooled or warmed as needed and pulled through the
Thermal Pads to achieve the pre-set patient target temperature . Patient temperature
can be controlled and monitored between 33°C and 37°C (9 .4ºF and 98 .6ºF) .
NOTE: The system will not go into Automatic Mode unless a patient’s temperature
probe is inserted into a patient and attached to the control unit.
CAUTION: The Control Unit will control and monitor core temperature using an
indwelling temperature probe attached to the system. Kimberly-Clark
recommends the use of a second site to verify core temperature while
controlling and monitoring temperature.
m
anual
m
Ode
:
The Manual Mode is functional with or without a patient temperature probe placed .
Water temperature is established and managed by the operator . Patient tempera-
ture is not controlled in Manual mode . The operator must monitor patient tempera-
ture when Manual Mode is active . Water temperature can be modified during the
Manual Mode Operation .
NOTE: Manual Mode can be used prior to placement of a temperature probe or
transition to Automatic Mode.
2 .3 KIMBERLY-CLARK*
Patient Warming
System M000
Control Unit User
Interfaces
2 .4 Modes of
Operation