* Patient Warming Systems
M000 Control Unit Operator’s Manual: U .S .
Tools Needed: 1/8" Hex Wrench
. Unpack both boxes associated with the KIMBERLY-CLARK* Patient Warming
System M000: the Control Unit (Part # 0000-0) and the Installation Kit
(Part # 00839-0) .
2 . Insert the Steel Pole into the moulded base of the cart . Align the depressions in
the pole with the screws in the base of the cart and tighten these screws with
a /8" Hex Wrench . (Figure )
Adjust the top pole such that the hook at the top
points away from the base .
3 . Ensure all four wheels of the cart won’t move by pressing down on the latch
on the wheel . This will aid in attaching the control unit to the cart .
4 . Remove the control unit from the plastic bag and wipe down any
condensation on the exterior of the unit .
5 . Remove the plugs from the Manifold on the back of the unit, and the foam
plug from the vent hole . (Figure 2)
6 . Open the pole clamp on the control unit by turning the handle on the side of
the unit counterclockwise . (Figure 3)
7 Carefully lift the control unit such that the clamp fits just on top of the
support ledge on the pole . Tighten down the pole clamp by turning the pole
clamp handle clockwise until just snug . (Figure 4) Raise top pole and rotate
so curve is extended over control unit .
8 . Remove the fill tube (Part # 00809-02) from its bag and attach it to the fill tube
port located just below the Fluid Delivery Line connection . (Figure 5)
9 . Remove the User Interface Cable [Part # 00832-0(42 in .) or Remote Display
Cable Part # 00832-02 (5 ft .)] from its bag and attach it to the UI Cable Port on
the back of the control unit .
N
ote
: It is important that the thumbscrews on the connector be tightened
securely to assure a good connection.
(Figures 6 & 7)
0 . Locate a butterfly nut and a cable clamp in the strain relief kit (Part # 00832-99) .
Place the UI Cable in the cable clamp (clamp should be just snug on the cable) .
Leaving approximately 3 in .(7 .5 cm) of cable from the clamp to the connector,
place the clamp on the top metal screw stud located next to the power outlet .
Secure the clamp with the butterfly nut provided .
Appendix A —
Installation & Test
Procedure
Figure 3
Figure 2
Figure 1
Control Unit
Support
Pole Clamp
Handle
Pole Clamp
Fluid Delivery
Line Port
(Manifold)