M000 Control Unit Operator’s Manual: U .S .
5 . When the machine is full, the pump will stop and the screen will read, “Fill
Complete .” Press Enter at this point .
6 . Gently shake and tap the Fill Tube to remove residual water from the line and
secure the line to the back of the unit using the pre-attached clamps .
7 . Attach the recirculating line (Part # 00809-04) to the end of the fluid deliver line .
8 . Press the Manual Key on the User Interface .
9 . The target temperature will begin blinking . Using the arrow keys, adjust the
temperature up or down to 4°C (05 .8ºF) . Press Enter when complete .
0 . Allow the machine to run for approximately 5 minutes . Verify that the flow rate
displayed is at least 3 L/min and that the temperature is climbing . The water
temperature should reach 4°C (05 .8ºF) within 0 minutes . Some overshoot
(approximately 0 .5°C) is normal, while the unit will eventually settle at
the target tempe/- 0 .°C/F .
. Change the water temperature set point to 30°C (86ºF) by pressing the Down
Arrow Key again until the screen reads, “Water Temp Target XX, press Enter to
change .” Adjust as before, and press Enter when complete .Verify that the
cooling fan comes on and that the water temperature begins to cool . The unit
should reach 30°C (86ºF) within 5 to 20 minutes .
2 . Once the unit has reached 30°C (86ºF), purge the system by pressing the Purge
Key . Verify that the water is removed from the recirculating line .
3 . Turn off the KIMBERLY-CLARK* Patient Warming System M000 Control Unit
and remove the recirculating line .
4 . The unit is now ready for use .
Appendix A —
Installation & Test
Procedure
(continued)
Figure 11
Figure 13
Figure 10
Figure 9
Figure 12
Latch
Patient
Temp 1 Port