3
K
imberly
-c
larK
* Patient Warming Systems
M000 Control Unit Operator’s Manual: U .S .
The Patient Temperature Trend Indicator reflects the change in the patient’s temper-
ature over the previous five minutes . Flashing Up arrows indicate patient tempera-
ture is increasing . Flashing Down arrows indicate patient temperature is decreasing .
The Patient Temperature Display indicates temperature from Patient Temperature
in degrees Centigrade (C) or degress Fahrenheit (F) . The display units can be
changed in the custom parameter menu found under Setup when in Stop Mode . The
temperature range that can be displayed is between 0°C and 42°C (50ºF to 08ºF) .
Water flows from the KIMBERLY-CLARK* Patient Warming System M000 Control
Unit to the thermal pads through the Fluid Delivery Line that connects to the back
of the control unit . The Fluid Delivery Line is reusable and should only be replaced
if defective (see Section 5 .2 Troubleshooting) . The Fluid Delivery Line comes in two
configurations accepting either 2 pads (Part # 00834-04) or 4 pads (Part # 00834-05) .
• Center circle no change or less than 0 .25ºC (0 .45ºF)
change per hour
• One arrow (up or down) – 0 .25ºC to 0 .5C (0 .45ºF to 0 .96ºF)
change per hour
• Two arrows (up or down) – 0 .5ºC to 0 .75ºC (0 .96ºF to .35ºF)
change per hour
• Three arrows (up or down) – 0 .75ºC to 2 .0ºC ( .35ºF to 3 .6ºF)
change per hour
• Three arrows flashing simultaneously – > 2 .0ºC (> 3 .6ºF)
change per hour
2 .6 Patient
Temperature
Trend Indicator
2 .7 Patient
Temperature
Display
2 .8 Fluid Delivery
Lines
Figure 4 Patient Temperature Trend Indicator
Fluid Delivery Line Port