TOS5300_INTERFACE
69
App
C
Default State
When the TOS5300 Series is initialized at the factory or when it receives an *RST or *RCL command,
its settings are set to the values shown in the following table.
When the power is turned on, the settings are set to the values that were in use immediately before
the POWER switch was turned off.
Item
Command
Setting after
initialization
Setting after command received
*RST
*RCL
Withstanding
voltage test
conditions
(ACW)
Measurement method
SENS:MODE
RMS
←
No change
Test voltage
SOUR:VOLT
0 V
←
Depends on the settings
in memory
Test voltage limit
SOUR:VOLT:PROT
5.5 kV
Start voltage on and off
SOUR:VOLT:STAR:STAT
OFF
Voltage rise time
SOUR:VOLT:SWE:TIM
0.1 s
Voltage fall time on and off
SOUR:VOLT:SWE:FALL:TIM:STAT
OFF
Upper limit
SENS:JUDG
0.02 mA
Lower limit
SENS:JUDG:LOW
0.01 mA
Lower limit on and off
SENS:JUDG:LOW:STAT
OFF
Test time
SOUR:VOLT:TIM
0.1 s
Test timer on and off
SOUR:VOLT:TIM:STAT
ON
Test voltage frequency
SOUR:VOLT:FREQ
50 Hz
Withstanding
voltage test
conditions
(DCW)
Test voltage
SOUR:DCW:VOLT
0 V
←
Depends on the settings
in memory
Test voltage limit
SOUR:DCW:VOLT:PROT
6.2 kV
Start voltage on and off
SOUR:DCW:VOLT:STAR:STAT
OFF
Voltage rise time
SOUR:DCW:VOLT:SWE:TIM
0.1 s
Upper limit
SENS:DCW:JUDG
0.02 mA
Lower limit
SENS:DCW:JUDG:LOW
0.01 mA
Lower limit on and off
SENS:DCW:JUDG:LOW:STAT
OFF
Test time
SOUR:DCW:VOLT:TIM
0.1 s
Test time on and off
SOUR:DCW:VOLT:STAT
ON
Judgment wait time
SENS:DCW:JUDG:DEL
0.1 s
Insulation
resistance test
conditions (IR)
Test voltage
SOUR:IR:VOLT
25 V
←
Depends on the settings
in memory
Test voltage limit
SOUR:IR:VOLT:PROT
1000 V
Upper limit
SENS:IR:JUDG
100 M
Ω
Upper limit on and off
SENS:IR:JUDG:STAT
OFF
Lower limit
SENS:IR:JUDG:LOW
1.00 M
Ω
Lower limit on and off
SENS:IR:JUDG:LOW:STAT
ON
Test time
SOUR:IR:VOLT:TIM
0.1 s
Test time on and off
SOUR:IR:VOLT:TIM:STAT
ON
Judgment wait time
SENS:IR:JUDG:DEL
0.1 s
Current detection response
speed
SENS:IR: MODE
Mid
Test mode
ACW voltage test/ DCW voltage
test/ Insulation resistance test
SOUR:FUNC:MODE
ACW
←
Depends on the settings
in memory
Содержание TOS5300
Страница 8: ...8 TOS5300_INTERFACE This page is intentionally blank ...
Страница 14: ...14 TOS5300_INTERFACE This page is intentionally blank ...
Страница 15: ...Command Reference This chapter explains topics such as command details and registers Remote Control ...
Страница 76: ...76 TOS5300_INTERFACE ...