64
TOS5300_INTERFACE
List of Messages
SYSTem subsystem
TEST subsystem
TRIGger subsystem
SCPI command
Setting
Default Response Effect
Description
R/W Nt
Program header
Parameter
Unit
SYST
:CONF
:BEEP:VOL
:FAIL
numeric
0.0 to 1.0
0.5
NR3
B
Sets the volume level of the buzzer
that is sounded when a FAIL judgment
occurs
R/W
3
:PASS
numeric
0.0 to 1.0
0.3
NR3
B
Sets the volume level of the buzzer
that is sounded when a PASS judgment
occurs
R/W
3
:CAL:DUE:CONT
numeric
3 to 36
month
12
NR1
B
Sets the calibration period
R/W
3
:DATE
NR1
2000 to 2099
NR1
Sets the date (year)
R/W
3
NR1
1 to 12
NR1
Sets the date (month)
NR1
1 to 31
NR1
Sets the date (day)
:PHOL
numeric
50m|100m|
200m|1|2|5|
INF
s
50m
NR3
B
Sets the PASS judgment result hold
time
R/W
3
:TIME
NR1
0 to 23
NR1
Sets the time (hour)
R/W
3
NR1
0 to 59
NR1
Sets the time (minutes)
NR1
0 to 59
NR1
Sets the time (seconds)
:ERR[:NEXT]
Queries error and event information
R
1
:KLOC
bool
OFF
NR1
B
Locks and unlocks panel operations
R/W
3
:LOC
Switches to local mode
W
2
:OPT
0 | 1
Queries options
R
3
:REM
Switches to remote mode; locks all
keys other than LOCAL and the STOP
switch
W
2
:RWL
Switches to remote mode; locks all
keys other than the STOP switch
W
2
:VERS
1999.0
Queries the version of the compliant
SCPI specifications
R
1
SCPI command
Setting
Default Response Effect
Description
R/W Nt
Program header
Parameter
Unit
TEST
:ABOR
Aborts testing (SEQ2)
W
3
:EXEC
Starts testing (starts the trigger function; SEQ2)
W
3
:PROT:CLE
Clears protection mode
W
3
SCPI command
Setting
Default Response Effect
Description
R/W Nt
Program header
Parameter
Unit
TRIG[:SEQ[1]] | TRIG[:ACQ] (sequence 1, ACQuire)
[:IMM]
Executes a software trigger (SEQ1)
W
1
:COUN
numeric
1 to 100
1
NR3
B
Executes a test with the applicable
trigger count (SEQ1)
R/W
3
:SOUR
char
IMM | BUS | TIM
| TEST
IMM
char
B
Sets the trigger source (SEQ1)
R/W
3
:TIM
numeric
1 to 60.0
S
0
NR3
B
Sets the trigger timer (SEQ1)
R/W
3
TRIG:SEQ2 | TRIG:TEST (sequence 2, TEST)
[:IMM]
Executes a software trigger (SEQ2)
W
1
:SOUR
char
IMM | BUS | EXT
IMM
char
B
Sets the trigger source (SEQ2)
R/W
3
Содержание TOS5300
Страница 8: ...8 TOS5300_INTERFACE This page is intentionally blank ...
Страница 14: ...14 TOS5300_INTERFACE This page is intentionally blank ...
Страница 15: ...Command Reference This chapter explains topics such as command details and registers Remote Control ...
Страница 76: ...76 TOS5300_INTERFACE ...