Test Execution (SEQuence2:TEST)
This function executes test. You have to specify the execution conditions in advance.
TRIG:SEQ2:SOUR / TRIG:TEST:SOUR
Sets the trigger source to apply to the SEQuence2 (TEST) group.
Command
TRIGger:SEQuence2:SOURce {IMMediate|BUS|EXTernal}
TRIGger:SEQuence2:SOURce?
TRIGger:TEST:SOURce {IMMediate|BUS|EXTernal}
TRIGger:TEST:SOURce?
Parameter
Value:
IMMediate
Begin testing immediately (default).
BUS
Wait for a software trigger—a *TRG, TRIG, or IEEE 488.1 get
(Group Execute Trigger) command—and then begin testing.
EXTernal
Begin testing when the START switch is pressed.
Response
Returns the trigger source in <character> format.
INIT:SEQ2 / INIT:NAME TEST / TEST:EXEC
Starts the test in the test mode that was set with SOUR:FUNC:MODE (starts the trigger function).
Use the trigger source that was set with TRIG:SEQ2:SOUR or TRIG:TEST:SOUR to start the test.
Command
INITiate[:IMMediate]:SEQuence2
INITiate[:IMMediate]:NAME TEST
TEST:EXECute
TRIG:SEQ2 / TRIG:TEST
Executes a software trigger for the SEQuence2 (TEST) group. This command is valid when
TRIG:SEQ2:SOUR or TRIG:TEST:SOUR is set to BUS.
Command
TRIGger:SEQuence2[:IMMediate]
TRIGger:TEST[:IMMediate]
TEST:ABOR
Aborts the SEQuence2 (TEST) group’s test. The measured data is discarded.
Command
TEST:ABORt
* RST