Insulation Resistance Test (IR) Conditions
These are commands for setting the insulation resistance test conditions.
Test voltage setting (IR)
SOUR:IR:VOLT
Sets the test voltage.
Command
SOURce:IR:VOLTage[:LEVel] {<numeric>|MIN|MAX}
SOURce:IR:VOLTage[:LEVel]? {MIN|MAX}
Parameter
Value:
25, 50, 100, 125, 250, 500, 1000 (The default is 25.)
Unit:
V
If you specify a voltage that cannot be used, the voltage is set to the next lowest value that can be
used.
Example
If you specify 999 V, the voltage is set to 500 V.
Response
Returns the test voltage setting in <NR3> format.
Limit voltage setting (IR)
SOUR:IR:VOLT:PROT
Sets the limit voltage.
Command
SOURce:IR:VOLTage:PROTection[:LEVel][:UPPer]
{<numeric>|MIN|MAX}
SOURce:IR:VOLTage:PROTection[:LEVel][:UPPer]? {MIN|MAX}
Parameter
Value:
25, 50, 100, 125, 250, 500, 1000 (The default is 1000.)
Unit:
V
Response
Returns the limit voltage in <NR3> format.