62
TOS5300_INTERFACE
List of Messages
SOURce subsystem
SCPI command
Setting
Default Response Effect
Description
R/W Nt
Program header
Parameter
Unit
SOUR
:FUNC:MODE
char
ACW | DCW | IR
ACW
char
A
Sets the test mode
R/W
3
[:ACW]:VOLT
:FREQ
numeric
50 | 60
HZ
50
NR3
A
Sets the test voltage frequency
R/W
3
[:LEV]
numeric
0 to 5.5 k
V
0
NR3
A
Sets the test voltage
R/W
3
:PROT[:LEV][:UPP]
numeric
0 to 5.5 k
V
5.5k
NR3
A
Sets the limit voltage
R/W
3
:STAR:STAT
bool
OFF
NR1
A
Turns the start voltage on and off
R/W
3
:SWE
:FALL:TIM:STAT
bool
OFF
NR1
A
Turns the voltage fall time on and off R/W
3
[:RISE]:TIM
numeric
0.1 to 10.0
S
0.1
NR3
A
Sets the voltage rise time
R/W
3
:TIM
numeric
0.1 to 999.0
S
0.1
NR3
A
Sets the test time
R/W
3
:STAT
bool
ON
NR1
A
Turns on or off the feature that ends
tests when their test time elapses
R/W
3
:DCW:VOLT
[:LEV]
numeric
0 to 6.2 k
V
0
NR3
A
Sets the test voltage
R/W
3
:PROT[:LEV][:UPP]
numeric
0 to 6.2 k
V
6.2 k
NR3
A
Sets the limit voltage
R/W
3
:STAR:STAT
bool
OFF
NR1
A
Turns the start voltage on and off
R/W
3
:SWE[:RISE]:TIM
numeric
0.1 to 10.0
S
0.1
NR3
A
Sets the voltage rise time
R/W
3
:TIM
numeric
0.1 to 999.0
S
0.1
NR3
A
Sets the test time
R/W
3
:STAT
bool
ON
NR1
A
Turns on or off the feature that ends
tests when their test time elapses
R/W
3
:IR:VOLT
[:LEV]
numeric
25 | 50 | 100 | 125
| 250 | 500 |1000
V
25
NR3
A
Sets the test voltage
R/W
3
:PROT[:LEV][:UPP]
numeric
25 | 50 | 100 | 125
| 250 | 500 |1000
V
1000
NR3
A
Sets the limit voltage
R/W
3
:TIM
numeric
0.1 to 999.0
S
0.1
NR3
A
Sets the test time
R/W
3
:STAT
bool
ON
NR1
A
Turns on or off the feature that ends
tests when their test time elapses
R/W
3
Содержание TOS5300
Страница 8: ...8 TOS5300_INTERFACE This page is intentionally blank ...
Страница 14: ...14 TOS5300_INTERFACE This page is intentionally blank ...
Страница 15: ...Command Reference This chapter explains topics such as command details and registers Remote Control ...
Страница 76: ...76 TOS5300_INTERFACE ...