50
TOS5300_INTERFACE
SCPI Register Model
OPERation status register (STATus:OPERation)
The OPERation status register is a 16-bit register that contains information about the normal
operating conditions of the TOS5300 Series.
STAT:OPER
Queries the event of the OPERation status register.
A query clears the contents of the register.
Command
STATus:OPERation[:EVENt]?
Response
Returns the event of the OPERation status register in <NR1> format.
STAT:OPER:COND
Queries the condition of the OPERation status register.
A query does not clear the contents of the register.
Command
STATus:OPERation:CONDition?
Response
Returns the condition of the OPERation status register in <NR1> format.
Bit
Bit weight
Bit name
Description
0
1
NOT USED
—
1
2
NOT USED
—
2
4
NOT USED
—
3
8
NOT USED
—
4
16
MEASuring (MEAS)
Measurements are being performed (SEQ1)
5
32
Waiting for TRIGger (WTG)
Waiting for triggers
6
64
NOT USED
—
7
128
NOT USED
—
8
256
PROTecting (PROT)
This bit is set to true when a bit is set in the
PROTecting status register.
9
512
High Voltage ON (HVON)
During testing or while voltage remains across
the output terminals
10
1024
TESTing (TEST)
This bit is set to true when a bit is set in the
TESTing status register.
11
2048
NOT USED
—
12
4096
NOT USED
—
13
8192
NOT USED
—
14
16384
PROGram is running (PROG)
During test execution (SEQ2)
15
32768
NOT USED
—